Biodegradable screw presents as a loose intra-articular body after anterior cruciate ligament reconstruction

被引:41
作者
MacDonald, P [1 ]
Arneja, S [1 ]
机构
[1] Univ Manitoba, Sect Orthoped, Winnipeg, MB R2H 3C3, Canada
关键词
anterior cruciate ligament; interference screw; complications; loose body; foreign body; revision arthroscopy;
D O I
10.1016/S0749-8063(03)00382-7
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report a case of intra-articular movement of a broken piece of a poly-L-lactide (PLLA) bioabsorbable interference screw from the femoral tunnel in anterior cruciate ligament (ACL) reconstruction with quadrupled semi-tendinosus and gracilis tendon grafts. Eleven months after initially successful ACL surgery, the patient felt a sudden locking of the knee without associated trauma or injury. The patient experienced pain and swelling episodes after heavy lifting with knee flexion at work, but without symptoms of giving way or locking. On revision arthroscopy, a broken part of a bioabsorbable interference screw was seen in the lateral compartment, which was subsequently removed without incident. The semitendinosus-gracilis graft appeared intact without disruption. After revision surgery, the patient's recovery was uneventful, with return to activity within a few months. This case further shows the problem of biointerference screw breakage in ACL reconstructive surgery and the need to bury the femoral bioabsorbable interference screw on graft fixation.
引用
收藏
页数:3
相关论文
共 8 条
[1]
Initial fixation strength of polylactic acid interference screws in anterior cruciate ligament reconstruction [J].
Abate, JA ;
Fadale, PD ;
Hulstyn, MJ ;
Walsh, WR .
ARTHROSCOPY, 1998, 14 (03) :278-284
[2]
An intra-articular bioabsorbable interference screw mimicking an acute meniscal tear 8 months after an anterior cruciate ligament reconstruction [J].
Bottoni, CR ;
DeBerardino, TM ;
Fester, EW ;
Mitchell, D ;
Penrod, BJ .
ARTHROSCOPY, 2000, 16 (04) :395-398
[3]
BUSH JC, 1998, AM J KNEE SURG, V11, P32
[4]
Biomechanical comparison between BioScrew and titanium alloy interference screws for bone-patellar tendon-bone graft fixation in anterior cruciate ligament reconstruction [J].
Caborn, DNM ;
Urban, WP ;
Johnson, DL ;
Nyland, J ;
Pienkowski, D .
ARTHROSCOPY, 1997, 13 (02) :229-232
[5]
Bioabsorbable polyglyconate interference screw fixation in anterior cruciate ligament reconstruction: A prospective computed tomography-controlled study [J].
Fink, C ;
Benedetto, KP ;
Hackl, W ;
Hoser, C ;
Freund, MC ;
Rieger, M .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2000, 16 (05) :491-498
[6]
Comparison of initial fixation strength between biodegradable and metallic interference screws and a press-fit fixation technique in a porcine model [J].
Seil, R ;
Rupp, S ;
Krauss, PW ;
Benz, A ;
Kohn, DM .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1998, 26 (06) :815-819
[7]
Biodegradable interference screw fixation exhibits pull-out force and stiffness similar to titanium screws [J].
Weiler, A ;
Windhagen, HJ ;
Raschke, MJ ;
Laumeyer, A ;
Hoffmann, RFG .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1998, 26 (01) :119-128
[8]
Secondary intra-articular dislocation of a broken bioabsorbable interference screw after anterior cruciate ligament reconstruction [J].
Werner, A ;
Wild, A ;
Ilg, A ;
Krauspe, R .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2002, 10 (01) :30-32