A 'mixed' self-assembled monolayer for an impedimetric immunosensor

被引:131
作者
Rickert, J
Gopel, W
Beck, W
Jung, G
Heiduschka, P
机构
[1] UNIV TUBINGEN,DEPT OPHTHALMOL,LAB THANOS,D-72076 TUBINGEN,GERMANY
[2] UNIV TUBINGEN,INST THEORET & PHYS CHEM,D-72076 TUBINGEN,GERMANY
[3] UNIV TUBINGEN,INST ORGAN CHEM,D-72076 TUBINGEN,GERMANY
关键词
immunosensing; impedance spectroscopy; peptide antigen; self-assembled monolayer; alkane thiols;
D O I
10.1016/0956-5663(96)85927-6
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A synthetic peptide with the amino acid sequence 135-154 of the capsid protein VP1 of the foot-and-mouth-disease virus was modified with omega-hydroxyundecanethiol and applied together with non-derivatised omega-hydroxyundecanethiol for consecutive adsorption onto gold electrodes according to self-assembling procedures. The binding of a specific antibody to prepared recognition layers could be monitored by measurement of impedance or capacitance. In order to avoid non-specific effects, all measurements were performed in the presence of BSA. The complex between the antigenic peptide and the antibody was split by applying 6 M urea solution. The gold electrodes were mounted into an optimised flow-through system in order to perform capacitance-time measurements. The immobilised peptide can be recognised repeatedly by specific antibodies. (C) 1996 Elsevier Science Limited
引用
收藏
页码:757 / 768
页数:12
相关论文