First human trials of a dry electrophysiology sensor using a carbon nanotube array interface

被引:88
作者
Ruffini, G. [1 ]
Dunne, S. [1 ]
Fuentemilla, L. [2 ]
Grau, C.
Farres, E. [1 ]
Marco-Pallares, J. [1 ,2 ]
Watts, P. C. P. [3 ]
Silva, S. R. P. [3 ]
机构
[1] Starlab, Barcelona 08035, Spain
[2] Univ Barcelona, Psychiat & Clin Psychobiol Dept, Neurodynam Lab, E-08007 Barcelona, Spain
[3] Univ Surrey, Adv Technol Inst, Nanoelect Ctr, Guildford, England
关键词
electrophysiology; EEG CNT nanotubes; electrodes dry;
D O I
10.1016/j.sna.2008.03.007
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We report the results from the first human trials of a new dry electrode sensor for surface biopotential applications. This sensor uses nanotechnology to improve performance and wearability. Standard electrodes for demanding applications such as EEG require skin preparation and/or the use of electrolytic substances to provide and effective electrical interface. The contact surface of the electrode described here is covered with an array of carbon nanotubes designed for penetration of the outer layers of the skin for improved electrical contact without the use of electrolytic gel. In previous papers we detailed the design concept and reported the results of the initial tests of a first prototype-immersion in saline solution and signal detection in animal skin. In this paper we describe the first human trials of the prototype, which indicate performance on par with state-of-the-art research-oriented wet electrodes. No side effects have been observed 6 months after the tests, and the subject did not report any pain or unusual sensations upon application of the electrode. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:275 / 279
页数:5
相关论文
共 11 条
[1]  
Altevogt B M., 2006, SLEEP DISORDERS SLEE
[2]  
*BIOS, BIOS ACT 2
[3]   Electrical conductivity of individual carbon nanotubes [J].
Ebbesen, TW ;
Lezec, HJ ;
Hiura, H ;
Bennett, JW ;
Ghaemi, HF ;
Thio, T .
NATURE, 1996, 382 (6586) :54-56
[4]   Investigation into the origin of the noise of surface electrodes [J].
Huigen, E ;
Peper, A ;
Grimbergen, CA .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2002, 40 (03) :332-338
[5]   Young's modulus of single-walled nanotubes [J].
Krishnan, A ;
Dujardin, E ;
Ebbesen, TW ;
Yianilos, PN ;
Treacy, MMJ .
PHYSICAL REVIEW B, 1998, 58 (20) :14013-14019
[6]  
NAATANEN R, 1983, ATTENTION BRAIN FUNC
[7]   Synthesis of large arrays of well-aligned carbon nanotubes on glass [J].
Ren, ZF ;
Huang, ZP ;
Xu, JW ;
Wang, JH ;
Bush, P ;
Siegal, MP ;
Provencio, PN .
SCIENCE, 1998, 282 (5391) :1105-1107
[8]  
Ruffini Giulio, 2006, Conf Proc IEEE Eng Med Biol Soc, V2006, P1826
[9]   A dry electrophysiology electrode using CNT arrays [J].
Ruffini, Giulio ;
Dunne, Stephen ;
Farres, Esteve ;
Marco-Pallares, Josep ;
Ray, Chris ;
Mendoza, Ernest ;
Silva, Ravi ;
Grau, Carles .
SENSORS AND ACTUATORS A-PHYSICAL, 2006, 132 (01) :34-41
[10]   HIGH-QUALITY RECORDING OF BIOELECTRIC EVENTS .2. LOW-NOISE, LOW-POWER MULTICHANNEL AMPLIFIER DESIGN [J].
VANRIJN, AC ;
PEPER, A ;
GRIMBERGEN, CA .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1991, 29 (04) :433-440