A Preliminary Biomechanical Assessment of a Polymer Composite Hip Implant Using an Infrared Thermography Technique Validated by Strain Gage Measurements

被引:16
作者
Bougherara, Habiba [1 ]
Rahim, Ehsan [1 ]
Shah, Suraj [1 ,2 ]
Dubov, Anton [2 ]
Schemitsch, Emil H. [2 ,3 ]
Zdero, Rad [1 ,2 ]
机构
[1] Ryerson Univ, Dept Mech & Ind Engn, Toronto, ON M5B 2K3, Canada
[2] St Michaels Hosp, Martin Orthopaed Biomech Lab, Toronto, ON M5B 1W8, Canada
[3] Univ Toronto, Fac Med, Dept Surg, Toronto, ON M5S 1A8, Canada
来源
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME | 2011年 / 133卷 / 07期
关键词
biomechanics; infrared thermography; stress; strain; composite hip implant; FEMORAL-SHAFT FRACTURES; FIXATION; DESIGN; STEMS;
D O I
10.1115/1.4004414
中图分类号
Q6 [生物物理学];
学科分类号
071011 [生物物理学];
摘要
With the resurgence of composite materials in orthopaedic applications, a rigorous assessment of stress is needed to predict any failure of bone-implant systems. For current biomechanics research, strain gage measurements are employed to experimentally validate finite element models, which then characterize stress in the bone and implant. Our preliminary study experimentally validates a relatively new nondestructive testing technique for orthopaedic implants. Lock-in infrared (IR) thermography validated with strain gage measurements was used to investigate the stress and strain patterns in a novel composite hip implant made of carbon fiber reinforced polyamide 12 (CF/PA12). The hip implant was instrumented with strain gages and mechanically tested using average axial cyclic forces of 840 N, 1500 N, and 2100 N with the implant at an adduction angle of 15 deg to simulate the single-legged stance phase of walking gait. Three-dimensional surface stress maps were also obtained using an IR thermography camera. Results showed almost perfect agreement of IR thermography versus strain gage data with a Pearson correlation of R-2 = 0.96 and a slope = 1.01 for the line of best fit. IR thermography detected hip implant peak stresses on the inferior-medial side just distal to the neck region of 31.14 MPa (at 840 N), 72.16 MPa (at 1500 N), and 119.86 MPa (at 2100 N). There was strong correlation between IR thermography-measured stresses and force application level at key locations on the implant along the medial (R-2 = 0.99) and lateral (R-2 = 0.83 to 0.99) surface, as well as at the peak stress point (R-2 = 0.81 to 0.97). This is the first study to experimentally validate and demonstrate the use of lock-in IR thermography to obtain three-dimensional stress fields of an orthopaedic device manufactured from a composite material. [DOI: 10.1115/1.4004414]
引用
收藏
页数:6
相关论文
共 28 条
[1]
Akay M, 1996, J BIOMED MATER RES, V31, P167
[2]
HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1993, 26 (08) :969-990
[3]
Twenty-five-year survivorship of two thousand consecutive primary Charnley total hip replacements - Factors affecting survivorship of acetabular and femoral components [J].
Berry, DJ ;
Harmsen, WS ;
Cabanela, ME ;
Morrey, BF .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (02) :171-177
[4]
The biomechanics of a validated finite element model of stress shielding in a novel hybrid total knee replacement [J].
Bougherara, H. ;
Zdero, R. ;
Mahboob, Z. ;
Dubov, A. ;
Shah, S. ;
Schemitsch, E. H. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2010, 224 (H10) :1209-1219
[5]
The biomechanics of the T2 femoral nailing system: a comparison of synthetic femurs with finite element analysis [J].
Bougherara, H. ;
Zdero, R. ;
Miric, M. ;
Shah, S. ;
Hardisty, M. ;
Zalzal, P. ;
Schemitsch, E. H. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2009, 223 (H3) :303-314
[6]
Design of a biomimetic polymer-composite hip prosthesis [J].
Bougherara, Habiba ;
Bureau, Martin ;
Campbell, Melissa ;
Vadean, Aurelian ;
Yahia, L'Hocine .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2007, 82A (01) :27-40
[7]
A preliminary biomechanical study of a novel carbon-fibre hip implant versus standard metallic hip implants [J].
Bougherara, Habiba ;
Zdero, Rad ;
Dubov, Anton ;
Shah, Suraj ;
Khurshid, Shaheen ;
Schemitsch, Emil H. .
MEDICAL ENGINEERING & PHYSICS, 2011, 33 (01) :121-128
[8]
A biomechanical assessment of modular and monoblock revision hip implants using FE analysis and strain gage measurements [J].
Bougherara, Habiba ;
Zdero, Rad ;
Shah, Suraj ;
Miric, Milan ;
Papini, Marcello ;
Zalzal, Paul ;
Schemitsch, Emil H. .
JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2010, 5
[9]
CF/PA12 composite femoral stems: Manufacturing and properties [J].
Campbell, Melissa ;
Denault, Johanne ;
Yahia, Lhocine ;
Bureau, Martin N. .
COMPOSITES PART A-APPLIED SCIENCE AND MANUFACTURING, 2008, 39 (05) :796-804
[10]
Finite element analysis of a femoral retrograde intramedullary nail subject to gait loading [J].
Cheung, G ;
Zalzal, P ;
Bhandari, M ;
Spelt, JK ;
Papini, M .
MEDICAL ENGINEERING & PHYSICS, 2004, 26 (02) :93-108