Flow injection-capillary electrophoresis system with contactless conductivity detection and hydrostatic pressure generated flow.: Application to the quantitative analysis of inorganic anions in water samples

被引:41
作者
Kubán, P
Kubán, P
Kubán, V
机构
[1] Texas Tech Univ, Dept Chem & Biochem, Lubbock, TX 79409 USA
[2] Mendel Univ Agr & Forestry, Dept Chem & Biochem, Brno, Czech Republic
关键词
calibration methods; capillary electrophoresis; contactless conductivity detection; flow injection analysis; inorganic anions; quantitative analysis;
D O I
10.1002/elps.200305433
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A simple and inexpensive flow injection-capillary electrophoresis (FI-CE) system with contactless conductivity detection (CCD) for automated quantitative analysis of chloride, nitrate, and sulfate in various water samples is demonstrated. A glass bottle containing the background electrolyte that is raised above the FI-CE interface generates a pulse-free, highly reproducible flow of the electrolyte through the FI-CE interface. The system operates at a flow rate of 300 muLmin(-1) with an injection volume of only 4 muL. The repeatability of peak areas (n = 18) was better than 0.81% RSD and the sample throughput was 90 samples per hour using the background electrolyte containing 12 MM L-histidine adjusted to pH 4.00 with acetic acid. The limits of detection were better than 125 mugL(-1) and were comparable to those obtained by conventional CE systems with CCD. Various calibration methods for FI-CE system with electrokinetic injection were tested and their suitability for the analysis of anions in real samples was evaluated.
引用
收藏
页码:1935 / 1943
页数:9
相关论文
共 35 条
[1]   On-line ion-exchange preconcentration in a flow injection system coupled to capillary electrophoresis for the direct determination of UV absorbing anions [J].
Arce, L ;
Kuban, P ;
Ríos, A ;
Valcárcel, M ;
Karlberg, B .
ANALYTICA CHIMICA ACTA, 1999, 390 (1-3) :39-44
[2]   Combination of flow injection with capillary electrophoresis. Part 5. Automated preconcentration and determination of pseudoephedrine in human plasma [J].
Chen, HW ;
Fang, ZL .
ANALYTICA CHIMICA ACTA, 1999, 394 (01) :13-22
[3]   Combination of flow injection with capillary electrophoresis. Part 3. On-line sorption column preconcentration capillary electrophoresis system [J].
Chen, HW ;
Fang, ZL .
ANALYTICA CHIMICA ACTA, 1997, 355 (2-3) :135-143
[4]   Combination of flow injection with capillary electrophoresis - Part 4. Automated multicomponent monitoring of drug dissolution [J].
Chen, HW ;
Fang, ZL .
ANALYTICA CHIMICA ACTA, 1998, 376 (02) :209-220
[5]  
da Silva JAF, 1998, ANAL CHEM, V70, P4339
[6]   INTERNAL STANDARDIZATION TECHNIQUE FOR CAPILLARY ZONE ELECTROPHORESIS [J].
DOSE, EV ;
GUIOCHON, GA .
ANALYTICAL CHEMISTRY, 1991, 63 (11) :1154-1158
[7]   On-line coupling of a miniaturized bioreactor with capillary electrophoresis, via a membrane interface, for monitoring the production of organic acids by microorganisms [J].
Ehala, S ;
Vassiljeva, I ;
Kuldvee, R ;
Vilu, R ;
Kaljurand, M .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2001, 371 (02) :168-173
[8]   Sequential injection sample introduction microfluidic-chip based capillary electrophoresis system [J].
Fang, Q ;
Wang, FR ;
Wang, SL ;
Liu, SS ;
Xu, SK ;
Fang, ZL .
ANALYTICA CHIMICA ACTA, 1999, 390 (1-3) :27-37
[9]   Combination of flow injection with capillary electrophoresis .1. The basic system [J].
Fang, ZL ;
Liu, ZS ;
Shen, Q .
ANALYTICA CHIMICA ACTA, 1997, 346 (02) :135-143
[10]   Combination of flow injection with capillary electrophoresis - Part 7. Microchip capillary electrophoresis system with flow injection sample introduction and amperometric detection [J].
Fu, CG ;
Fang, ZL .
ANALYTICA CHIMICA ACTA, 2000, 422 (01) :71-79