Defects in ProTaper S1 instruments after clinical use: fractographic examination

被引:132
作者
Cheung, GSP
Peng, B
Bian, Z
Shen, Y
Darvell, BW
机构
[1] Wuhan Univ, Key Lab Oral Biomed Engn, Minist Educ, Dept Operat Dent & Endodont,Sch Stomatol, Wuhan 430072, Peoples R China
[2] Univ Hong Kong, Fac Dent, Hong Kong, Peoples R China
关键词
fatigue; fracture; nickel-titanium; rotary instrument; shear;
D O I
10.1111/j.1365-2591.2005.01020.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Aim To investigate the mode of failure of a brand of nickel-titanium instruments separated during clinical use, by detailed examination of the fracture surface. Methodology A total of 122 ProTaper S1 instruments were discarded from an endodontic clinic at a stomatological school in China over a period of 17 months; 28 had fractured. These fractured instruments were ultrasonically cleaned, autoclaved and then examined under a scanning electron microscope. From the lateral view the fracture was classified into 'torsional' or 'flexural'. The specimens were then re-mounted and the presence of characteristics of shear failure and fatigue striations was recorded under high-power view of the fracture surface. The difference in the mean lengths of fractured segment between the shear and fatigue groups was compared using Student's t-test. Results Twenty-seven separated instruments were available for analysis. Under low-power magnification, only two fell into the category of 'torsional' failure when examined laterally; the others appeared to be 'flexural'. Close examination of the fracture surface revealed the presence of fatigue striations in 18 specimens. Nine instruments (including the two putative 'torsional' failures above) fell into the shear fracture group, in which fatigue striations were absent or characteristics of shear failure of the material were found. The mean length of fractured segments resulting from fatigue failure (4.3 +/- 1.9 mm) was significantly greater than that for shear failure (2.5 +/- 0.8 mm) (P < 0.001, two-sample t-test). Conclusions Examination of the fracture surface at high magnification is essential to reveal features that may indicate the possible origin of cracks and the mode of material failure. Macroscopic or lateral examination of separated instruments would fail to reveal the true mechanism of failure. Fatigue seems to be an important reason for the separation of rotary instruments during clinical use.
引用
收藏
页码:802 / 809
页数:8
相关论文
共 37 条
[1]   K3 endo, ProTaper, and ProFile systems: Breakage and distortion, in severely curved roots of molars [J].
Ankrum, MT ;
Hartwell, GR ;
Truitt, JE .
JOURNAL OF ENDODONTICS, 2004, 30 (04) :234-237
[2]   Evaluation of single-use rotary nickel-titanium instruments [J].
Arens, FC ;
Hoen, MM ;
Steiman, HR ;
Dietz, GC .
JOURNAL OF ENDODONTICS, 2003, 29 (10) :664-666
[3]  
ASM International, 1987, ASM HDB, V12
[4]   Comparative analysis of torsional and bending stresses in two mathematical models of nickel-titanium rotary instruments: ProTaper versus ProFile [J].
Berutti, E ;
Chiandussi, G ;
Gaviglio, I ;
Ibba, A .
JOURNAL OF ENDODONTICS, 2003, 29 (01) :15-19
[5]   Analysis of forces developed during mechanical preparation of extracted teeth using Profile NiTi rotary instruments [J].
Blum, JY ;
Cohen, A ;
Machtou, P ;
Micallef, JP .
INTERNATIONAL ENDODONTIC JOURNAL, 1999, 32 (01) :24-31
[6]   Effect of surface treatment of NiTi alloy on its corrosion behavior in Hanks' solution [J].
Cisse, O ;
Savadogo, O ;
Wu, M ;
Yahia, L .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 2002, 61 (03) :339-345
[7]  
COLLINS JA, 1993, FAILURE MAT MECH DES, P1
[8]   Comparison of fracture rate, deformation rate, and efficiency between rotary endodontic instruments driven at 150 rpm and 350 rpm [J].
Daugherty, DW ;
Gound, TG ;
Comer, TL .
JOURNAL OF ENDODONTICS, 2001, 27 (02) :93-95
[9]  
DAUSKARDT RH, 1989, P MRS INT M ADV MAT, V9, P243
[10]  
EWALDS HL, 1989, FRACTURE MECH 5, P226