Target margins for random geometrical treatment uncertainties in conformal radiotherapy

被引:101
作者
Bel, A [1 ]
vanHerk, M [1 ]
Lebesque, JV [1 ]
机构
[1] NETHERLANDS CANC INST,ANTONI VAN LEEUWENHOEK HUIS,NL-1066 CX AMSTERDAM,NETHERLANDS
关键词
3-D treatment planning; geometrical treatment uncertainties; target margins; morphological image processing;
D O I
10.1118/1.597745
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this study we investigate a method for positioning the margin-required around the clinical target volume (CTV) to account for the random geometrical treatment uncertainties during conformal radiotherapy. These uncertainties are introduced by patient setup errors and CTV motion within the patient. Three-dimensional dose distributions are calculated for two four-field box techniques and a three-field technique, using rectangular fields. In addition, dose calculations are performed for four prostate cases, treated with a three-field conformal technique. The effects of random rotational and translational deviations on the delivered dose are described as a convolution of the ''static'' dose with the distribution of the deviations. For the rectangular field techniques, these convolutions are performed with a range of standard deviations (SDs) of the distribution of random translations (0-7 mm in the three directions) and rotations (0 degrees-5 degrees around the main axes). Two centers of rotation are considered: the isocenter and a position that is 3.5 cm shifted with respect to the isocenter. For the prostate cases, the random deviations are estimated by combining the results from organ motion and setup accuracy studies. The required margin is defined as the change in the position of the static 95% isodose surface by the convolution and it is approximated by a morphological erosion operator, applied to the static 95% isodose surface. When the center of rotation coincides with the isocenter the change in the position of the static 95% isodose surface can accurately be described by an erosion operator. For the rectangular field techniques, the margin is equal to about 0.7 SD of the distribution of translations, independent of the distribution of rotations. When the center of rotation does not coincide with the isocenter and rotations are considerable, the margin is strongly place dependent, and the accuracy of the approximation by an erosion operator is much lower. In conclusion, margins for random uncertainties can be approximated by a dilation operator (inverse of an erosion operator) when the center of rotational deviations coincides with the isocenter. The size of the margin is about 0.7 SD of the distribution of translations. When rotational deviations are present and the center of rotation does not coincide with the isocenter, the margin can become strongly place dependent and the convolution computation should be incorporated in the planning system. (C) 1996 American Association of Physicists in Medicine.
引用
收藏
页码:1537 / 1545
页数:9
相关论文
共 23 条
  • [1] [Anonymous], 1993, PRESCR REC REP PHOT
  • [2] MEASUREMENT OF PROSTATE MOVEMENT OVER THE COURSE OF ROUTINE RADIOTHERAPY USING IMPLANTED MARKERS
    BALTER, JM
    SANDLER, HM
    LAM, K
    BREE, RL
    LICHTER, AS
    TENHAKEN, RK
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 31 (01): : 113 - 118
  • [3] ANALYSIS OF PROSTATE AND SEMINAL-VESICLE MOTION
    BEARD, CJ
    BUSSIERE, MR
    PLUNKETT, ME
    COLEMAN, CN
    KIJEWSKI, PK
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 27 : 136 - 136
  • [4] High-precision prostate cancer irradiation by clinical application of an offline patient setup verification procedure, using portal imaging
    Bel, A
    Vos, PH
    Rodrigus, PTR
    Creutzberg, CL
    Visser, AG
    Stroom, JC
    Lebesque, JV
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 35 (02): : 321 - 332
  • [5] MAXIMIZING SETUP ACCURACY USING PORTAL IMAGES AS APPLIED TO A CONFORMAL BOOST TECHNIQUE FOR PROSTATIC-CANCER
    BIJHOLD, J
    LEBESQUE, JV
    HART, AAM
    VIJLBRIEF, RE
    [J]. RADIOTHERAPY AND ONCOLOGY, 1992, 24 (04) : 261 - 271
  • [6] ROUTINE CLINICAL ONLINE PORTAL IMAGING FOLLOWED BY IMMEDIATE FIELD ADJUSTMENT USING A TELE-CONTROLLED PATIENT COUCH
    DENEVE, W
    VANDENHEUVEL, F
    DEBEUKELEER, M
    COGHE, M
    THON, L
    DEROOVER, P
    VANLANCKER, M
    STORME, G
    [J]. RADIOTHERAPY AND ONCOLOGY, 1992, 24 (01) : 45 - 54
  • [7] DAILY MONITORING AND CORRECTION OF RADIATION-FIELD PLACEMENT USING A VIDEO-BASED PORTAL IMAGING-SYSTEM - A PILOT-STUDY
    EZZ, A
    MUNRO, P
    PORTER, AT
    BATTISTA, J
    JAFFRAY, DA
    FENSTER, A
    OSBORNE, S
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1992, 22 (01): : 159 - 165
  • [8] A RANDOMIZED TRIAL OF PATIENT REPOSITIONING DURING RADIOTHERAPY USING A MEGAVOLTAGE IMAGING-SYSTEM
    GILDERSLEVE, J
    DEARNALEY, DP
    EVANS, PM
    LAW, M
    RAWLINGS, C
    SWINDELL, W
    [J]. RADIOTHERAPY AND ONCOLOGY, 1994, 31 (02) : 161 - 168
  • [9] IMAGE-ANALYSIS USING MATHEMATICAL MORPHOLOGY
    HARALICK, RM
    STERNBERG, SR
    ZHUANG, XH
    [J]. IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1987, 9 (04) : 532 - 550
  • [10] HOEKSTRA CJM, 1992, RADIOTHER ONCOL, V24, P45