Multinight "Bedside" Closed-Loop Control for Patients with Type 1 Diabetes

被引:42
作者
Brown, Sue A. [1 ]
Kovatchev, Boris P. [1 ]
Breton, Marc D. [1 ]
Anderson, Stacey M. [1 ]
Keith-Hynes, Patrick [1 ]
Patek, Stephen D. [1 ]
Jiang, Boyi [1 ]
Ben Brahim, Najib [1 ]
Vereshchetin, Paul [1 ]
Bruttomesso, Daniela [2 ]
Avogaro, Angelo [2 ]
Del Favero, Simone [2 ]
Boscari, Federico [2 ]
Galasso, Silvia [2 ]
Visentin, Roberto [2 ]
Monaro, Marco [2 ]
Cobelli, Claudio [2 ]
机构
[1] Univ Virginia, Ctr Diabet Technol, Charlottesville, VA USA
[2] Univ Padua, Padua, Italy
基金
美国国家卫生研究院;
关键词
ARTIFICIAL PANCREAS; INSULIN DELIVERY; GLUCOSE CONTROL; ADULTS;
D O I
10.1089/dia.2014.0259
中图分类号
R5 [内科学];
学科分类号
100201 [内科学];
摘要
Background: Studies of closed-loop control (CLC) systems have improved glucose levels in patients with type 1 diabetes. In this study we test a new CLC concept aiming to "reset" the patient overnight to near-normoglycemia each morning, for several consecutive nights. Subjects and Methods: Ten insulin pump users with type 1 diabetes (mean age, 46.4 +/- 8.5 years) were enrolled in a two-center (in the United States and Italy) randomized crossover trial comparing 5 consecutive nights of CLC (23:00-07:00 h) in an outpatient setting versus sensor-augmented insulin pump therapy of the same duration at home. Primary end points included time spent in 80-140 mg/dL as measured by continuous glucose monitoring overnight and fasting blood glucose distribution at 7:00 h. Results: Compared with sensor-augmented pump therapy, CLC improved significantly time spent between 80 and 140 mg/dL (54.5% vs. 32.2%; P<0.001) and between 70 and 180 mg/dL (85.4% vs. 59.1%; P<0.001); CLC reduced the mean glucose level at 07:00 h (119.3 vs. 152.9 mg/dL; P<0.001) and overnight mean glucose level (139.0 vs. 170.3 mg/dL; P<0.001) using a marginally lower amount of insulin (6.1 vs. 6.8 units; P=0.1). Tighter overnight control led to improved daytime control on the next day: the overnight/next-day control correlation was r=0.52, PConclusions: Multinight CLC of insulin delivery (artificial pancreas) results in significant improvement in morning and overnight glucose levels and time in target range, with the potential to improve daytime control when glucose levels were "reset" to near-normoglycemia each morning.
引用
收藏
页码:203 / 209
页数:7
相关论文
共 17 条
[1]
MD-Logic Artificial Pancreas System A pilot study in adults with type 1 diabetes [J].
Atlas, Eran ;
Nimri, Revital ;
Miller, Shahar ;
Grunberg, Eli A. ;
Phillip, Moshe .
DIABETES CARE, 2010, 33 (05) :1072-1076
[2]
Fully Integrated Artificial Pancreas in Type 1 Diabetes: Modular Closed-Loop Glucose Control Maintains Near Normoglycemia [J].
Breton, Marc ;
Farret, Anne ;
Bruttomesso, Daniela ;
Anderson, Stacey ;
Magni, Lalo ;
Patek, Stephen ;
Man, Chiara Dalla ;
Place, Jerome ;
Demartini, Susan ;
Del Favero, Simone ;
Toffanin, Chiara ;
Hughes-Karvetski, Colleen ;
Dassau, Eyal ;
Zisser, Howard ;
Doyle, Francis J., III ;
De Nicolao, Giuseppe ;
Avogaro, Angelo ;
Cobelli, Claudio ;
Renard, Eric ;
Kovatchev, Boris .
DIABETES, 2012, 61 (09) :2230-2237
[3]
MEDICINE A Pancreas in a Box [J].
Clery, Daniel .
SCIENCE, 2014, 343 (6167) :133-135
[4]
Pilot Studies of Wearable Outpatient Artificial Pancreas in Type 1 Diabetes [J].
Cobelli, Claudio ;
Renard, Eric ;
Kovatchev, Boris P. ;
Keith-Hynes, Patrick ;
Ben Brahim, Najib ;
Place, Jerome ;
Del Favero, Simone ;
Breton, Marc ;
Farret, Anne ;
Bruttomesso, Daniela ;
Dassau, Eyal ;
Zisser, Howard ;
Doyle, Francis J., III ;
Patek, Stephen D. ;
Avogaro, Angelo .
DIABETES CARE, 2012, 35 (09) :E65-E67
[5]
Artificial Pancreas Past, Present, Future [J].
Cobelli, Claudio ;
Renard, Eric ;
Kovatchev, Boris .
DIABETES, 2011, 60 (11) :2672-2682
[6]
Managed by machine [J].
Dolgin, Elie .
NATURE, 2012, 485 (7398) :S6-S8
[7]
A Bihormonal Closed-Loop Artificial Pancreas for Type 1 Diabetes [J].
El-Khatib, Firas H. ;
Russell, Steven J. ;
Nathan, David M. ;
Sutherlin, Robert G. ;
Damiano, Edward R. .
SCIENCE TRANSLATIONAL MEDICINE, 2010, 2 (27) :27ra27
[8]
Overnight Closed- Loop Insulin Delivery in Young People With Type 1 Diabetes: A Free- Living, Randomized Clinical Trial [J].
Hovorka, Roman ;
Elleri, Daniela ;
Thabit, Hood ;
Allen, Janet M. ;
Leelarathna, Lalantha ;
El-Khairi, Ranna ;
Kumareswaran, Kavita ;
Caldwell, Karen ;
Calhoun, Peter ;
Kollman, Craig ;
Murphy, Helen R. ;
Acerini, Carlo L. ;
Wilinska, Malgorzata E. ;
Nodale, Marianna ;
Dunger, David B. .
DIABETES CARE, 2014, 37 (05) :1204-1211
[9]
Manual closed-loop insulin delivery in children and adolescents with type 1 diabetes: a phase 2 randomised crossover trial [J].
Hovorka, Roman ;
Allen, Janet M. ;
Elleri, Daniela ;
Chassin, Ludovic J. ;
Harris, Julie ;
Xing, Dongyuan ;
Kollman, Craig ;
Hovorka, Tomas ;
Larsen, Anne Mette F. ;
Nodale, Marianna ;
De Palma, Alessandro ;
Wilinska, Malgorzata E. ;
Acerini, Carlo L. ;
Dunger, David B. .
LANCET, 2010, 375 (9716) :743-751
[10]
Hughes Colleen S, 2010, J Diabetes Sci Technol, V4, P1146