Do SpaceLabs ambulatory non-invasive blood pressure recorders measure blood pressure consistently over several years use?

被引:13
作者
Amoore, JN
Dewar, D
Gough, K
Padfield, PL
机构
[1] Royal Infirm Edinburgh NHS Trust, Dept Phys Med, Edinburgh EH16 4SA, Midlothian, Scotland
[2] Western Gen Hosp, Cardiovasc Med Clin, Edinburgh EH4 2XU, Midlothian, Scotland
[3] Western Gen Hosp, Dept Med, Edinburgh EH4 2XU, Midlothian, Scotland
关键词
ambulatory blood pressure measurement equipment; measurement consistency; maintenance;
D O I
10.1097/00126097-200502000-00009
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective To assess the measurement consistency of SpaceLabs ambulatory recorders (Spacelabs, Washington, USA) that are in regular use. Methods A total of 14 SpaceLabs 90207 and one 90217 ambulatory recorders were tested for measurement consistency using the Dynatech Cuff Link (Dynatech, Nevada, USA), a commercially available non-invasive blood pressure (NIBP) simulator. The NIBP recorders were tested at a range of pressures with 20 repeated determinations at a simulated 120/80 mmHg and five repeated determinations at simulated pressures of 80/50, 100/80, 150/100, 200/165 and 250/195 mmHg. Tests were carried out in 1998, 2002 and 250/195 mmHg. Tests were carried out in 1998, 2002 and late 2003 or early 2004. Conclusions All 15 SpaceLabs recorders measured consistently over the 6 years with 89.5% of the differences in average pressures, recorded by any particular device at each recorded pressure, less than 2 mmHg between successive test episodes. The maximum difference was 4.5 mmHg and 60.1% of the differences were less than 1 mmHg. The measurements for all devices were within the tolerances specified by the supplier for the device when tested with the simulator. Maintenance records also show that most devices required breakdown maintenance less than once every 3 years. The results show that the SpaceLabs devices maintain measurement consistency in the demanding conditions of ambulatory pressure recording over several years. Blood Press Monit 10:51-56 (C) 2005 Lippincott Williams Wilkins.
引用
收藏
页码:51 / 56
页数:6
相关论文
共 16 条
[1]  
Amoore J N, 1997, J Clin Eng, V22, P93
[2]  
Amoore JN, 2000, BLOOD PRESS MONIT, V5, P69
[3]   Evaluation of the Critikon 8100 and Spacelabs 90207 non-invasive blood pressure monitors using a test simulator [J].
Amoore, JN ;
Geake, WB .
JOURNAL OF HUMAN HYPERTENSION, 1997, 11 (03) :163-169
[4]  
Amoore JN, 2000, BLOOD PRESS MONIT, V5, P81
[5]  
Association for the Advancement of Medical Instrumentation, 1993, ANSI AAMI SP, P40
[6]  
Baumgart, 1998, Blood Press Monit, V3, P303
[7]   An automated system for the functional evaluation of oscillometric non-invasive blood pressure monitors [J].
Geake, WB ;
Amoore, JN ;
Scott, DHT .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1995, 19 (05) :162-176
[8]   Limitations of current validation protocols for home blood pressure monitors for individual patients [J].
Gerin, W ;
Schwartz, AR ;
Schwartz, JE ;
Pickering, TG ;
Davidson, KW ;
Bress, J ;
O'Brien, E ;
Atkins, N .
BLOOD PRESSURE MONITORING, 2002, 7 (06) :313-318
[9]   Sphygmomanometers in use in general practice: an overlooked aspect of quality in patient care [J].
Knight, T ;
Leech, F ;
Jones, A ;
Walker, L ;
Wickramasinghe, R ;
Angris, S ;
Rolfe, P .
JOURNAL OF HUMAN HYPERTENSION, 2001, 15 (10) :681-684
[10]   The mercury sphygmomanometer should be abandoned before it is proscribed [J].
Markandu, ND ;
Whitcher, F ;
Arnold, A ;
Carney, C .
JOURNAL OF HUMAN HYPERTENSION, 2000, 14 (01) :31-36