Determination of knee joint cartilage thickness using three-dimensional magnetic resonance chondro-crassometry (3D MR-CCM)

被引:130
作者
Eckstein, F
Gavazzeni, A
Sittek, H
Haubner, M
Losch, A
Milz, S
Englmeier, KH
Schulte, E
Putz, R
Reiser, M
机构
[1] UNIV MUNICH, KLINIKUM GROSSHADERN, INST RADIOL DIAGNOST, D-8000 MUNICH, GERMANY
[2] GSF FORSCHUNGSZENTRUM UNWELT & GESUNDHEIT, INST MED INFORMAT & SYST FORSCH, MUNICH, GERMANY
[3] UNIV MAINZ, INST ANAT, D-6500 MAINZ, GERMANY
关键词
cartilage; cartilage thickness; knee; MR; magnetic resonance (MR) technology;
D O I
10.1002/mrm.1910360213
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The objective of this article was to analyze the accuracy and precision with which the quantitative distribution of articular cartilage can be determined in the knee joint using MRI. A three-dimensional (3D) technique that accounts for the out-of-plane deviation of the interface normal in strongly curved joint surfaces (3D MR-CCM) has been developed for cartilage thickness measurements. Eight cadaveric knee-joint specimens and six volunteers were imaged using a fat-suppressed gradient-echo sequence at a resolution of 2 x 0.31 x 0.31 mm(3). Cartilage volumes and topographical thickness maps were obtained and compared with those derived from anatomical sections by image analysis. The deviation of the MR volumes from those of the sections was 1-12%, the coefficient of variation after repositioning ranged from 2.9% (patella) to 8.2% (lateral tibial plateau). Between 60% and 80% of all image points could be attributed to identical thickness intervals, less than 20% deviating by more than 0.5 mm. The intraobserver and interobserver reproducibilities were very high in both the specimens and the volunteers. In the knee joint, 3D reconstructions of the cartilages, and measurements that take into account the out-of-plane deviation of the interface normals (3D MR-CCM), are required.
引用
收藏
页码:256 / 265
页数:10
相关论文
共 35 条
[1]   MEASUREMENT IN MEDICINE - THE ANALYSIS OF METHOD COMPARISON STUDIES [J].
ALTMAN, DG ;
BLAND, JM .
JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES D-THE STATISTICIAN, 1983, 32 (03) :307-317
[2]   QUANTITATION OF ARTICULAR SURFACE-TOPOGRAPHY AND CARTILAGE THICKNESS IN KNEE JOINTS USING STEREOPHOTOGRAMMETRY [J].
ATESHIAN, GA ;
SOSLOWSKY, LJ ;
MOW, VC .
JOURNAL OF BIOMECHANICS, 1991, 24 (08) :761-776
[3]   AN ASYMPTOTIC SOLUTION FOR THE CONTACT OF 2 BIPHASIC CARTILAGE LAYERS [J].
ATESHIAN, GA ;
LAI, WM ;
ZHU, WB ;
MOW, VC .
JOURNAL OF BIOMECHANICS, 1994, 27 (11) :1347-1360
[4]   A THEORETICAL SOLUTION FOR THE FRICTIONLESS ROLLING-CONTACT OF CYLINDRICAL BIPHASIC ARTICULAR-CARTILAGE LAYERS [J].
ATESHIAN, GA ;
WANG, HQ .
JOURNAL OF BIOMECHANICS, 1995, 28 (11) :1341-1355
[5]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[6]   DETECTION OF KNEE HYALINE CARTILAGE DEFECTS USING FAT-SUPPRESSED 3-DIMENSIONAL SPOILED GRADIENT-ECHO MR-IMAGING - COMPARISON WITH STANDARD MR-IMAGING AND CORRELATION WITH ARTHROSCOPY [J].
DISLER, DG ;
MCCAULEY, TR ;
WIRTH, CR ;
FUCHS, MD .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 165 (02) :377-382
[7]   FAT-SUPPRESSED SPOILED GRASS IMAGING OF KNEE HYALINE CARTILAGE - TECHNIQUE OPTIMIZATION AND COMPARISON WITH CONVENTIONAL MR-IMAGING [J].
DISLER, DG ;
PETERS, TL ;
MUSCOREIL, SJ ;
RATNER, LM ;
WAGLE, WA ;
COUSINS, JP ;
RIFKIN, MD .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1994, 163 (04) :887-892
[8]   THE MORPHOLOGY OF ARTICULAR-CARTILAGE ASSESSED BY MAGNETIC-RESONANCE-IMAGING (MRI) - REPRODUCIBILITY AND ANATOMICAL CORRELATION [J].
ECKSTEIN, F ;
SITTEK, H ;
MILZ, S ;
PUTZ, R ;
REISER, M .
SURGICAL AND RADIOLOGIC ANATOMY, 1994, 16 (04) :429-438
[9]   Magnetic resonance chondro-crassometry (MR CCM): A method for accurate determination of articular cartilage thickness [J].
Eckstein, F ;
Sittek, H ;
Gavazzeni, A ;
Schulte, E ;
Milz, S ;
Kiefer, B ;
Reiser, M ;
Putz, R .
MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (01) :89-96
[10]  
ECKSTEIN F, 1995, RADIOLOGE, V35, P87