European Concerted Action on Anticoagulation. Quality assessment of the CoaguChek mini and TAS PT-NC point-of-care whole-blood prothrombin time monitors

被引:17
作者
Poller, L [1 ]
Keown, M
Chauhan, N
van den Besselaar, AMHP
Tripodi, A
Shiach, C
Jespersen, J
机构
[1] Univ Manchester, Sch Biol Sci, ECAA Cent Facil, Manchester M13 9PT, Lancs, England
[2] Leiden Univ, Med Ctr, Haemostasis & Thrombosis Res Ctr, Leiden, Netherlands
[3] Univ Milan, IRCCS, Maggiore Hosp, A Bianchi Bonomi Hemophilia & Thrombosis Ctr, Milan, Italy
[4] Manchester Royal Infirm, Dept Haematol, Manchester M13 9WL, Lancs, England
[5] Univ So Denmark, Dept Thrombosis Res, Esbjerg, Denmark
[6] Ribe Cty Hosp, Dept Clin Biochem, Esbjerg, Denmark
关键词
D O I
10.1373/clinchem.2003.019653
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: International Normalized Ratios (INRs) for prothrombin time obtained with the CoaguChek Mini and TAS (RapidPointCoag) PT-NC systems are markedly different and also differ from the "true" INR. There is therefore a need for local quality assessment (QA) of the two systems. Methods: A set of 60 lyophilized artificially depleted and 60 lyophilized coumarin plasmas were tested at 10 centers on both point-of-care testing monitors. Subsets of three and five plasmas were selected as QA plasmas and compared with the remaining 55 to assess the relative ability of the systems to characterize performance at the individual centers. The incidence of aberrant results (outliers; >15% deviation from the true INR) was also recorded. The expected incidence with the QA plasmas was calculated and compared. Results: On both systems, INR with the common sets of 55 lyophilized plasmas varied considerably between centers. With the TAS PT-NC, subsets of five and three European Concerted Action on Anticoagulation (ECAA) artificially depleted plasmas gave good correlation with the 55 plasmas, but the coumarin plasmas performed less well. With the CoaguChek Mini, correlation was good with sets of five artificially depleted QA plasmas and reasonable with three but was less satisfactory with the coumarin plasmas. Outliers were detected with both types of plasmas on both test systems but with variable success. Conclusions: With the TAS PT-NC, three ECAA artificially depleted lyophilized plasmas provided reliable QA, but five lyophilized coumarin plasmas were required. With the CoaguChek Mini, five artificially depleted plasmas gave reliable QA but coumarin plasmas gave poorer results. ECAA QA plasmas provide a local system for checking INRs obtained with monitors of both types. (C) 2004 American Association for Clinical Chemistry.
引用
收藏
页码:537 / 544
页数:8
相关论文
共 18 条
[1]   OPTIMAL ORAL ANTICOAGULANT-THERAPY IN PATIENTS WITH MECHANICAL HEART-VALVES [J].
CANNEGIETER, SC ;
ROSENDAAL, FR ;
WINTZEN, AR ;
VANDERMEER, FJM ;
VANDENBROUCKE, JP ;
BRIET, E .
NEW ENGLAND JOURNAL OF MEDICINE, 1995, 333 (01) :11-17
[2]   European Concerted Action on Anticoagulation. Correction of displayed international normalized ratio on two point-of-care test whole-blood prothrombin time monitors (CoaguChek Mini and TAS PT-NC) by independent international sensitivity index calibration [J].
Poller, L ;
Keown, M ;
Chauhan, N ;
van den Besselaar, AMHP ;
Tripodi, A ;
Shiach, C ;
Jespersen, J .
BRITISH JOURNAL OF HAEMATOLOGY, 2003, 122 (06) :944-949
[3]   European Concerted Action on Anticoagulation (ECAA): an assessment of a method for ISI calibration of two whole blood point-of-care PT monitor systems based on lyophilized plasmas using whole blood equivalent PT [J].
Poller, L ;
Keown, M ;
Chauhan, N ;
Van Den Besselaar, AMHP ;
Tripodi, A ;
Shiach, C ;
Jespersen, J .
JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2003, 1 (04) :766-772
[4]   Reliability of international normalised ratios from two point of care test systems: comparison with conventional methods [J].
Poller, L ;
Keown, M ;
Chauhan, N ;
van den Besselaar, AMHP ;
Tripodi, A ;
Shiach, C ;
Jespersen, J .
BMJ-BRITISH MEDICAL JOURNAL, 2003, 327 (7405) :30-32A
[5]   European Concerted Action on Anticoagulation - Minimum numbers of lyophilized plasma samples for ISI calibration of CoaguChek and TAS point-of-care whole blood prothrombin time monitors [J].
Poller, L ;
Keown, M ;
Chauhan, N ;
van den Besselaar, AMHP ;
Tripodi, A ;
Jespersen, J ;
Shiach, C .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2003, 119 (02) :232-240
[6]   The importance of "like to like" ISI calibrations with freeze dried plasmas [J].
Poller, L ;
van den Besselaar, AMHP ;
Jespersen, J ;
Tripodi, A ;
Houghton, D .
JOURNAL OF CLINICAL PATHOLOGY, 1998, 51 (04) :275-279
[7]   European Concerted Action on Anticoagulation (ECAA). An assessment of lyophilised plasmas for ISI calibration of CoaguChek and TAS whole blood prothrombin time monitors [J].
Poller, L ;
Keown, M ;
Chauhan, N ;
van den Besselaar, AMHP ;
Tripodi, A ;
Shiach, C ;
Jespersen, J .
JOURNAL OF CLINICAL PATHOLOGY, 2003, 56 (02) :114-119
[8]  
Poller L, 2002, THROMB HAEMOSTASIS, V88, P992
[9]   European Concerted Action on Anticoagulation (ECAA). Minimum number of centres for reliable International Sensitivity Index calibration of CoaguChek and TAS point-of-care whole blood monitors [J].
Poller, L ;
Keown, M ;
Chauhan, N ;
van den Besselaar, AMHP ;
Tripodi, A ;
Shiach, C ;
Jespersen, J .
THROMBOSIS RESEARCH, 2002, 107 (1-2) :61-66
[10]  
Poller L, 1996, THROMB HAEMOSTASIS, V76, P977