On the shared control of an EMG-controlled prosthetic hand: Analysis of user-prosthesis interaction

被引:316
作者
Cipriani, Christian [1 ,2 ]
Zaccone, Franco [1 ]
Micera, Silvestro [1 ]
Carrozza, M. Chiara [1 ]
机构
[1] Scuola Super Sant Anna, Adv Robot Technol & Syst Lab, I-56025 Pontedera, Italy
[2] IMT Lucca Inst Adv Studies, I-55100 Lucca, Italy
关键词
biorobotics; electromyographic (EMG) classification; prosthetic hand; shared control; vibrotactile feedback;
D O I
10.1109/TRO.2007.910708
中图分类号
TP24 [机器人技术];
学科分类号
080202 ; 1405 ;
摘要
An anthropomorphic underactuated prosthetic hand, endowed with position and force sensors and controlled by means of myoelectric commands, is used to perform experiments of hierarchical shared control. Three different hierarchical control strategies combined with a vibrotactile feedback system have been developed and tested by able-bodied subjects through grasping tasks used in activities of daily living (ADLs). The first goal is to find a good tradeoff between good grasping capabilities and low attention required by the user to complete grasping tasks, without addressing advanced algorithm for electromyographic processing. The second goal is to understand whether a vibrotactile feedback system is subjectively or objectively useful and how it changes users' performance. Experiments showed that users were able to successfully operate the device in the three control strategies, and that the grasp success increased with more interactive control. Practice has proven that when too much effort is required, subjects do not do their best, preferring, instead, a less-interactive control strategy. Moreover, the experiments showed that when grasping tasks are performed under visual control, the enhanced proprioception offered by a vibrotactile system is practically not exploited. Nevertheless, in subjective opinion, feedback seems to be quite important.
引用
收藏
页码:170 / 184
页数:15
相关论文
共 43 条
[31]  
*OTT BOCK, 2008, OTT BOCK HEALTH
[32]   Schema design and implementation of the grasp-related mirror neuron system [J].
Oztop, E ;
Arbib, MA .
BIOLOGICAL CYBERNETICS, 2002, 87 (02) :116-140
[33]  
PYLATIUK C, P 2004 IEEE MEBS INT, V2, P4262
[34]   COGNITIVE FEEDBACK FOR USE WITH FES UPPER EXTREMITY NEUROPROSTHESES [J].
RISO, RR ;
IGNAGNI, AR ;
KEITH, MW .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1991, 38 (01) :29-38
[35]  
RODRIGUEZCHEU LE, 2005, P IEEE RSJ INT C INT, P541
[36]  
ROESCHLEIN RA, 1989, PROSTHET ORTHOT INT, V13, P14
[37]  
Salisbury JK Jr, 1985, ROBOT HANDS MECH MAN
[38]  
SASAKI Y, P 2002 2 JOINT ENG M, V3, P2402
[39]   A hydraulically driven multifunctional prosthetic hand [J].
Schulz, S ;
Pylatiuk, C ;
Reischl, M ;
Martin, J ;
Mikut, R ;
Bretthauer, G .
ROBOTICA, 2005, 23 :293-299
[40]  
SCHULZ S, 2001, P INT C ROB AUT SEOU, P2347