Increasing the sensitivity of Listeria monocytogenes assays:: evaluation using ELISA and amperometric detection

被引:36
作者
Crowley, EL [1 ]
O'Sullivan, CK [1 ]
Guilbault, GG [1 ]
机构
[1] Natl Univ Ireland Univ Coll Cork, Lab Sensor Dev, Cork, Ireland
关键词
D O I
10.1039/a806875e
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An immunosensor for the detection of Listeria monocytogenes was developed. ELISA and amperometric studies were run in parallel to develop a more sensitive and rapid assay for the bacterium. Conditions for the immunosensor were primarily characterised using ELISA. A direct sandwich assay was employed and the affinities of two polyclonal (goat and rabbit) and one monoclonal (mouse) anti-L. monocytogenes antibodies were compared using this format. Owing to low sensitivity being obtained with all antibodies, biotin-avidin amplification and an indirect sandwich assay were employed. The system was then transferred to screen-printed electrodes (SPEs), the primary antibody being immobilised by cross-linking with 1-ethyl-3-(3-dimethylaminopropyl)-carbodiimide, and the mode of detection being amperometric. Various parameters (limit of detection, working range, incubation time, cross-reactivity) of the systems were characterised. The effect of direct incubation in milk is also discussed. The final immunosensor had a working range of 1 x 10(6)-1 x 10(3) cells ml(-1) and a detection limit of 9 x 10(2) cells ml(-1). The assay took about 3.5 h to complete.
引用
收藏
页码:295 / 299
页数:5
相关论文
共 15 条
[1]  
Adam MR, 1995, FOOD MICROBIOL, P187
[2]  
CURIALE MS, 1994, J AOAC INT, V77, P1472
[3]  
Feldsine PT, 1997, J AOAC INT, V80, P775
[4]  
Feldsine PT, 1997, J AOAC INT, V80, P791
[5]   A NOVEL LIPOSOME-BASED ELECTROCHEMICAL BIOSENSOR FOR THE DETECTION OF HEMOLYTIC MICROORGANISMS [J].
KIM, HJ ;
BENNETTO, HP ;
HALABLAB, MA .
BIOTECHNOLOGY TECHNIQUES, 1995, 9 (06) :389-394
[6]  
MARTIN A, 1993, J AOAC INT, V76, P632
[7]   A quartz crystal microbalance displacement assay for Listeria monocytogenes [J].
Minunni, M ;
Mascini, M ;
Carter, RM ;
Jacobs, MB ;
Lubrano, GJ ;
Guilbault, GG .
ANALYTICA CHIMICA ACTA, 1996, 325 (03) :169-174
[8]  
PATEL PD, 1994, RAPID ANAL TECHNIQUE, P63
[9]   DEVELOPMENT OF A PIEZOELECTRIC IMMUNOSENSOR FOR THE DETECTION OF SALMONELLA-TYPHIMURIUM [J].
PRUSAKSOCHACZEWSKI, E ;
LUONG, JHT ;
GUILBAULT, GG .
ENZYME AND MICROBIAL TECHNOLOGY, 1990, 12 (03) :173-177
[10]   ALKALINE-PHOSPHATASE AS A LABEL FOR A HETEROGENEOUS IMMUNOELECTROCHEMICAL SENSOR - AN ELECTROCHEMICAL STUDY [J].
ROSEN, I ;
RISHPON, J .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1989, 258 (01) :27-39