A mathematical model for the assessment of hemodynamic parameters using quantitative contrast echocardiography

被引:25
作者
Chen, XC
Schwarz, KQ
Phillips, D
Steinmetz, SD
Schlief, R
机构
[1] Univ Rochester, Dept Med, Rochester, NY 14642 USA
[2] Univ Rochester, Rochester Ctr Biomed Ultrasound, Rochester, NY 14642 USA
[3] Univ Rochester, Dept Elect Engn, Rochester, NY 14642 USA
[4] Schering AG, Dept Clin Dev Magnet Resonance & Ultrasound Contr, D-1000 Berlin, Germany
关键词
attenuation; contrast agent; hemodynamics; scattering;
D O I
10.1109/10.678610
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A mathematical model for the assessment of hemodynamic parameters using quantitative echocardiography is presented. The method involves the intravenous injection of are ultrasonic echo contrast agent. The relative enhancement of the backscattered ultrasound intensity is measured as a function of time (the time-intensity curve), From this measurement, the volume flow rate (cardiac output) and the mixing volume are calculated. Relevant acoustic properties of the ultrasound contrast agent are discussed, An in vitro experiment is performed to corroborate the theory presented.
引用
收藏
页码:754 / 765
页数:12
相关论文
共 14 条
[1]  
BLEEKER H, 1990, J ULTRAS MED, V9, P461
[2]   NORMALIZATION OF ULTRASONIC SCATTERING MEASUREMENTS TO OBTAIN AVERAGE DIFFERENTIAL SCATTERING CROSS-SECTIONS FOR TISSUES [J].
CAMPBELL, JA ;
WAAG, RC .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1983, 74 (02) :393-399
[3]   The measurement of backscatter coefficient from a broadband pulse-echo system: A new formulation [J].
Chen, XC ;
Phillips, D ;
Schwarz, KQ ;
Mottley, JG ;
Parker, KJ .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1997, 44 (02) :515-525
[4]   CARDIAC-OUTPUT MEASUREMENTS - A REVIEW OF CURRENT TECHNIQUES AND RESEARCH [J].
EHLERS, KC ;
MYLREA, KC ;
WATERSON, CK ;
CALKINS, JM .
ANNALS OF BIOMEDICAL ENGINEERING, 1986, 14 (03) :219-239
[5]  
FEINSTEIN SB, 1991, CARDIAC IMAGING
[6]   METHOD OF DATA REDUCTION FOR ACCURATE DETERMINATION OF ACOUSTIC BACKSCATTER COEFFICIENTS [J].
MADSEN, EL ;
INSANA, MF ;
ZAGZEBSKI, JA .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1984, 76 (03) :913-923
[7]  
MORSE PM, 1986, THEORETICAL ACOUSTIC, pCH8
[8]  
SCHWARZ K, 1995, ULTRASOUND MED BIOL, P231
[9]   The Doppler kinetics of microbubble echo contrast [J].
Schwarz, KQ ;
Chen, X ;
Bezante, GP ;
Phillips, D ;
Schlief, R .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1996, 22 (04) :453-462
[10]   QUANTITATIVE ECHO CONTRAST CONCENTRATION MEASUREMENT BY DOPPLER SONOGRAPHY [J].
SCHWARZ, KQ ;
BEZANTE, GP ;
CHEN, XC ;
SCHLIEF, R .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1993, 19 (04) :289-297