Estimation of changes in cardiac output from the arterial blood pressure waveform in the upper limb

被引:92
作者
Linton, NWF [1 ]
Linton, RAF [1 ]
机构
[1] St Thomas Hosp, Rayne Inst, London SE1 7EH, England
关键词
heart; cardiac output; arterial pressure; measurement techniques; pulse contour analysis; surgery; cardiovascular; monitoring; cardiopulmonary;
D O I
10.1093/bja/86.4.486
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We have developed a new pulse contour cardiac output (PulseCO) algorithm based on frequency analysis studies of the arterial system. PulseCO was compared with thermodilution cardiac output (TDCO) in IO patients undergoing cardiac surgery. Results from one patient were unsuitable for analysis. In the remaining nine patients, 142 TDCO determinations were compared with PulseCO after logarithmic transformation and after being normalized by the initial cardiac output in each patient. Each determination was usually the average of three measurements. Least squares regression gave y=0.77x (r(2)=0.81) and the limits of agreement were from -26% to +21%. The accuracy of PulseCO in determining short-term changes in cardiac output was assessed by comparing the ratios of consecutive PulseCO determinations with the ratios of the corresponding, consecutive TDCO determinations. Least squares regression gave y=0.71x (r(2)=0.70) and the limits of agreement were from -21% to +25%. After phenylephrine had been given to five patients, PulseCO showed an increase in systemic vascular resistance consistent with the known pharmacological actions of the drug. The PulseCO algorithm was incorporated into a computer program that acquires arterial pressure data from an analogue-to-digital converter and displays beat-to-beat trend values.
引用
收藏
页码:486 / 496
页数:11
相关论文
共 29 条
[11]   AN ANALYSIS OF THE RELATIONSHIP BETWEEN CENTRAL AORTIC AND PERIPHERAL UPPER LIMB PRESSURE WAVES IN MAN [J].
KARAMANOGLU, M ;
OROURKE, MF ;
AVOLIO, AP ;
KELLY, RP .
EUROPEAN HEART JOURNAL, 1993, 14 (02) :160-167
[12]   PRESSURE WAVE-PROPAGATION IN A MULTIBRANCHED MODEL OF THE HUMAN UPPER-LIMB [J].
KARAMANOGLU, M ;
GALLAGHER, DE ;
AVOLIO, AP ;
OROURKE, MF .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 1995, 269 (04) :H1363-H1369
[13]   FUNCTIONAL ORIGIN OF REFLECTED PRESSURE WAVES IN A MULTIBRANCHED MODEL OF THE HUMAN ARTERIAL SYSTEM [J].
KARAMANOGLU, M ;
GALLAGHER, DE ;
AVOLIO, AP ;
OROURKE, MF .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 1994, 267 (05) :H1681-H1688
[14]   ESTIMATION OF STROKE VOLUME IN DOG BY A PULSE CONTOUR METHOD [J].
KOUCHOUKOS, NT ;
SHEPPARD, LC ;
MCDONALD, DA .
CIRCULATION RESEARCH, 1970, 26 (05) :611-+
[15]   THE STATIC ELASTIC PROPERTIES OF 45 HUMAN THORACIC AND 20 ABDOMINAL AORTAS INVITRO AND THE PARAMETERS OF A NEW MODEL [J].
LANGEWOUTERS, GJ ;
WESSELING, KH ;
GOEDHARD, WJA .
JOURNAL OF BIOMECHANICS, 1984, 17 (06) :425-435
[16]   THE PRESSURE DEPENDENT DYNAMIC ELASTICITY OF 35 THORACIC AND 16 ABDOMINAL HUMAN AORTAS INVITRO DESCRIBED BY A 5-COMPONENT MODEL [J].
LANGEWOUTERS, GJ ;
WESSELING, KH ;
GOEDHARD, WJA .
JOURNAL OF BIOMECHANICS, 1985, 18 (08) :613-620
[17]   EFFECT OF NITROGLYCERIN ON AORTIC IMPEDANCE, DIAMETER, AND PULSE-WAVE VELOCITY [J].
LATSON, TW ;
HUNTER, WC ;
KATOH, N ;
SAGAWA, K .
CIRCULATION RESEARCH, 1988, 62 (05) :884-890
[18]   Lithium dilution cardiac output measurement: A comparison with thermodilution [J].
Linton, R ;
Band, D ;
OBrien, T ;
Jonas, M ;
Leach, R .
CRITICAL CARE MEDICINE, 1997, 25 (11) :1796-1800
[19]   A multicenter evaluation of a new continuous cardiac output pulmonary artery catheter system [J].
Mihm, FG ;
Gettinger, A ;
Hanson, CW ;
Gilbert, HC ;
Stover, EP ;
Vender, JS ;
Beerle, B ;
Haddow, G .
CRITICAL CARE MEDICINE, 1998, 26 (08) :1346-1350
[20]  
NICHOLS WW, 1973, BIOMED ENG, V8, P376