Determination of cadmium with a portable, battery-powered tungsten coil atomic absorption spectrometer

被引:46
作者
Batchelor, JD [1 ]
Thomas, SE [1 ]
Jones, BT [1 ]
机构
[1] Wake Forest Univ, Dept Chem, Winston Salem, NC 27109 USA
关键词
tungsten coil; cadmium; atomic absorption spectrometry; urine; soil; charge-coupled device; near-line background correction;
D O I
10.1366/0003702981944986
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A portable atomic absorption spectrometer is described that is powered by a 12 V car battery. The atomization device is a tungsten coil extracted from a common projector lamp bulb. The coil is housed in a glass cell, and the atomization environment is purged with 10% H-2 in Ar (available as a standard commercial welding gas) to prevent oxidation. The detection system is a miniature charge-coupled device (CCD) spectrometer mounted on a PC board. The device is controlled by a laptop computer. Background correction measurements are performed by the near-line method. Aqueous standard solutions provide a Cd detection limit of 3 mu g/L (60 pg) for a 20 mu L sample volume. Accuracies for the determination of Cd in urine, soil, and stream water are typically better than 90%, with relative standard deviations in the 10% range.
引用
收藏
页码:1086 / 1091
页数:6
相关论文
共 15 条
[1]  
*ATSDR, 1991, TOX PROF CADM
[2]   SIMPLE LOW-COST TUNGSTEN-COIL ATOMIZER FOR ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY [J].
BERNDT, H ;
SCHALDACH, G .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1988, 3 (05) :709-712
[3]   ANALYTICAL EVALUATION OF A TUNGSTEN COIL ATOMIZER FOR CADMIUM, LEAD, CHROMIUM, MANGANESE, NICKEL AND COBALT DETERMINATION BY ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY [J].
BRUHN, CG ;
AMBIADO, FE ;
CID, HJ ;
WOERNER, R ;
TAPIA, J ;
GARCIA, R .
ANALYTICA CHIMICA ACTA, 1995, 306 (2-3) :183-192
[4]  
GARTRELL MJ, 1986, J ASSOC OFF ANA CHEM, V69, P146
[5]   DETERMINATION OF CADMIUM IN BIOLOGICAL-MATERIALS BY TUNGSTEN COIL ATOMIC-ABSORPTION SPECTROMETRY [J].
GINE, MF ;
KRUG, FJ ;
SASS, VA ;
REIS, BF ;
NOBREGA, JA ;
BERNDT, H .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (02) :243-245
[6]  
GUNDERSON EL, 1988, J ASSOC OFF ANA CHEM, V71, P1200
[7]  
Ingle Jr J. D., 1988, SPECTROCHEMICAL ANAL
[8]   Determination of lead in blood by tungsten coil electrothermal atomic absorption spectrometry [J].
Krug, FJ ;
Silva, MM ;
Oliveira, PV ;
Nobrega, JA .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1995, 50 (12) :1469-1474
[9]   APPLICATION OF ELECTRODEPOSITION TECHNIQUES TO FLAMELESS ATOMIC-ABSORPTION SPECTROMETRY .1. DETERMINATION OF CADMIUM WITH A TUNGSTEN FILAMENT [J].
LUND, W ;
LARSEN, BV .
ANALYTICA CHIMICA ACTA, 1974, 70 (02) :299-310
[10]   A low-cost tungsten filament atomizer for measuring lead in blood by atomic absorption spectrometry [J].
Parsons, PJ ;
Qiao, HC ;
Aldous, KM ;
Mills, E ;
Slavin, W .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1995, 50 (12) :1475-1480