Preparation and characterization of monolithic porous capillary columns loaded with chromatographic particles

被引:168
作者
Dulay, MT [1 ]
Kulkarni, RP [1 ]
Zare, RN [1 ]
机构
[1] Stanford Univ, Dept Chem, Stanford, CA 94305 USA
关键词
D O I
10.1021/ac9806456
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Using sol-gel technology, a porous glass matrix (xerogel) is formed in a capillary column and acts as a support for a stationary phase of chromatographic particles used in capillary electrochromatography. Preparation of the sol-gel matrix and immobilization of the octadecylsilica (ODS) stationary phase occur in a single step. The presence of the particles in the column greatly reduces matrix cracking caused by internal pressure differentials within the pores of the sol-gel matrix. Good electroosmotic now is achieved in part because of the inherent negative charge of both the particles and the sol-gel matrix. The performance of these sol-gel/ODS capillary columns was evaluated with a mixture of aromatic and nonaromatic organic compounds. Efficiencies of up to 80 000 plates/m were observed in columns with immobilized 3-mu m ODS particles. The efficiency and resolution are enhanced when 3-mu m ODS particles are used in place of the 5-mu m particles.
引用
收藏
页码:5103 / 5107
页数:5
相关论文
共 44 条
  • [41] WEINBERGER R, 1993, PRACTICAL CAPILIARY
  • [42] CAPILLARY ELECTROCHROMATOGRAPHY - ANALYSIS OF POLYCYCLIC AROMATIC-HYDROCARBONS
    YAN, C
    DADOO, R
    ZHAO, H
    ZARE, RN
    RAKESTRAW, DJ
    [J]. ANALYTICAL CHEMISTRY, 1995, 67 (13) : 2026 - 2029
  • [43] CRITICAL STRESS INTENSITY FACTORS OF WET GELS
    ZARZYCKI, J
    [J]. JOURNAL OF NON-CRYSTALLINE SOLIDS, 1988, 100 (1-3) : 359 - 363
  • [44] SYNTHESIS OF GLASSES FROM GELS - THE PROBLEM OF MONOLITHIC GELS
    ZARZYCKI, J
    PRASSAS, M
    PHALIPPOU, J
    [J]. JOURNAL OF MATERIALS SCIENCE, 1982, 17 (11) : 3371 - 3379