Short and long term variability of the interrupter technique under field and standardised conditions in 3-6 year old children

被引:27
作者
Beelen, RMJ
Smit, HA
van Strien, RT
Koopman, LP
Brussee, JE
Brunekreef, B
Gerritsen, J
Merkus, PJFM
机构
[1] Natl Inst Publ Hlth & Environm, Dept Chron Dis Epidemiol, NL-3720 BA Bilthoven, Netherlands
[2] Univ Utrecht, Dept Environm & Occupat Hlth, Inst Risk Assessment Sci, Utrecht, Netherlands
[3] Sophia Childrens Univ Hosp, Dept Pediat, Div Resp Med, Erasmus Med Ctr, Rotterdam, Netherlands
[4] Univ Groningen Hosp, Dept Pediat Pulmonol, Groningen, Netherlands
关键词
D O I
10.1136/thorax.58.9.761
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Background: The short and long term variability of the interrupter technique was assessed to determine whether interrupter resistance is a stable individual characteristic over time. The effect of field and standardised measurement conditions on the within-subject variability of the interrupter technique was also examined. Methods: The interrupter technique was studied under field and standardised conditions in children aged 3-6 years. Under field conditions, five investigators performed the measurements using two different measurement devices in random sequence. Both short term (20-30 minutes) and long term variability (median 38 days) were assessed in 32 children. Under standardised conditions, a single investigator conducted all measurements using a single device; the repeated measurements were conducted at the same time of day in a familiar quiet classroom. Long term variability (median 11 days) was estimated in 15 children. Within-subject standard deviations were estimated by analysis of variance with adjustment for the effects of different investigators and measurement devices on within-subject variability under field conditions. Results: Under field conditions within-subject standard deviations for short and long term variability were 0.10 kPa/l/s (adjusted 0.10 kPa/l/s) and 0.13 kPa/l/s (adjusted 0.14 kPa/l/s), respectively. Under standardised conditions the within-subject standard deviation for long term variability was 0.10 kPa/l/s. Conclusions: Measurement of interrupter resistance under field conditions only slightly increased the within-subject variability compared with standardised conditions. The results indicate that interrupter resistance is a stable individual characteristic over a period of some weeks.
引用
收藏
页码:761 / 764
页数:4
相关论文
共 20 条
[1]  
BATES JHT, 1987, RESP PHYSIOL, V70, P301, DOI 10.1016/S0034-5687(87)80052-X
[2]   Lung function measurement in awake young children [J].
Bisgaard, H ;
Klug, B .
EUROPEAN RESPIRATORY JOURNAL, 1995, 8 (12) :2067-2075
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   A portable device based on the interrupter technique to measure bronchodilator response in schoolchildren [J].
Bridge, PD ;
Lee, H ;
Silverman, M .
EUROPEAN RESPIRATORY JOURNAL, 1996, 9 (07) :1368-1373
[5]   Measurement of airway resistance using the interrupter technique in preschool children in the ambulatory setting [J].
Bridge, PD ;
Ranganathan, S ;
McKenzie, SA .
EUROPEAN RESPIRATORY JOURNAL, 1999, 13 (04) :792-796
[6]   EVALUATION OF THE INTERRUPTER TECHNIQUE FOR THE USE OF ASSESSING AIRWAY-OBSTRUCTION IN CHILDREN [J].
CARTER, ER ;
STECENKO, AA ;
POLLOCK, BH ;
JAEGER, MJ .
PEDIATRIC PULMONOLOGY, 1994, 17 (04) :211-217
[7]   How should airways resistance be measured in young children: mask or mouthpiece? [J].
Child, F ;
Clayton, S ;
Davies, S ;
Fryer, AA ;
Jones, PW ;
Lenney, W .
EUROPEAN RESPIRATORY JOURNAL, 2001, 17 (06) :1244-1249
[8]   REPEATABILITY AND METHOD COMPARISON [J].
CHINN, S .
THORAX, 1991, 46 (06) :454-456
[9]  
Fleiss J., 1986, Reliability of measurement: the design and analysis of clinical experiments
[10]  
Kannisto S, 1999, PEDIATR PULM, V27, P203, DOI 10.1002/(SICI)1099-0496(199903)27:3<203::AID-PPUL9>3.0.CO