Respiratory Monitoring System on the Basis of Capacitive Textile Force Sensors

被引:55
作者
Hoffmann, T. [1 ,2 ]
Eilebrecht, B. [1 ,2 ]
Leonhardt, S. [1 ,2 ]
机构
[1] Rhein Westfal TH Aachen, Philips Chair Med Informat Technol, Aachen, Germany
[2] Chair Med Informat Technol, D-52074 Aachen, Germany
关键词
Capacitive force sensors; capacitive measurement; respiration monitoring; textile force sensors; INDUCTIVE PLETHYSMOGRAPH; TIDAL VOLUME; VENTILATION; EXERCISE;
D O I
10.1109/JSEN.2010.2082524
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A novel approach is presented for non-constrictive long-term-monitoring of respiration which could particularly become suitable for home care applications. The system is based on textile integrated force sensors which detect expansion of the thorax during respiration and allow wireless data transmission for maximum mobility. Possible applications include long-term monitoring of patients with chronic pulmonary diseases, early recognition of diseases and the performance measurement of athletes during exercise. Results from performance tests under various conditions are presented. The tests show a high correlation with measurements made with a flow meter as gold standard, allowing conclusions about the respiratory volume. Different respiration patterns are reliably detected and can easily be subdivided even during intensive motion of the person. The sensors' textile buildup may allow a future integration into clothing and thus a minimal constriction for the patient. Nevertheless, the respiratory volume estimations are still not accurate enough mainly because of the use of a simplified volume calculation model and the disregard of the abdominal volume changes.
引用
收藏
页码:1112 / 1119
页数:8
相关论文
共 18 条
[11]   Tidal volume and respiratory timing derived from a portable ventilation monitor [J].
McCool, FD ;
Wang, J ;
Ebi, KL .
CHEST, 2002, 122 (02) :684-691
[12]  
Meyer J, 2006, TENTH IEEE INTERNATIONAL SYMPOSIUM ON WEARABLE COMPUTERS, PROCEEDINGS, P69
[13]  
Paradiso Rita, 2004, Stud Health Technol Inform, V108, P253
[14]  
Popov T. A., 1996, Journal of Clinical Engineering, V21, P446
[15]  
VanLangenhove L, 2007, WOODHEAD PUBL SER TE, P1
[16]   The LifeShirt - An advanced system for ambulatory measurement of respiratory and cardiac function [J].
Wilhelm, FH ;
Roth, WT ;
Sackner, MA .
BEHAVIOR MODIFICATION, 2003, 27 (05) :671-691
[17]   Measurement of exercise ventilation by a portable respiratory inductive plethysmograph [J].
Witt, Jonathan D. ;
Fisher, Jason R. K. O. ;
Guenette, Jordan A. ;
Cheong, Krystie A. ;
Wilson, Brock J. ;
Sheel, A. William .
RESPIRATORY PHYSIOLOGY & NEUROBIOLOGY, 2006, 154 (03) :389-395
[18]  
Zhang J, 2001, J Tongji Med Univ, V21, P35