The development and test of a device for the reconstruction of 3-D position and orientation by means of a kinematic sensor assembly with rate gyroscopes and accelerometers

被引:92
作者
Giansanti, D [1 ]
Maccioni, G [1 ]
Macellari, V [1 ]
机构
[1] Ist Super Sanita, Dipartimento Tecnol & Salute, I-00161 Rome, Italy
关键词
accelerometer; accuracy; human movement analysis; simulation; 3-D rigid body position and orientation;
D O I
10.1109/TBME.2005.847404
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this paper, we propose a device for the Position and Orientation (P&O) reconstruction of human segmental locomotion tasks. It is based on three mono-axial accelerometers and three angular velocity sensors, geometrically arranged to form two orthogonal terns. The device was bench tested using step-by-step motor-based equipment. The characteristics of the six channels under bench test conditions were: crosstalk absent, non linearity <= +/- 0, 1% fs, hysteresis < 0, 1% fs, accuracy 0, 3% fs, overall resolution better than 0, 04 deg/s, 2 * g * 10(-4). The device was validated with the stereophotogrammetric body motion analyzer during the execution of three different locomotion tasks: stand-to-sit, sit-to-stand, gait-initiation. Results obtained comparing the trajectories of the two methods showed that the errors were lower than 3 *10(-2) M and 2 deg during a 4s of acquisition and lower than 6 *10(-3) m. and 0.2 deg during the effective duration of a locomotory task; showing that the wearable device hereby presented permits the 3-D reconstruction of the movement of the body segment to which it is affixed for time-limited clinical applications.
引用
收藏
页码:1271 / 1277
页数:7
相关论文
共 22 条
[1]   BIOMECHANICAL INVESTIGATION OF WRIST KINEMATICS [J].
ANDREWS, JG ;
YOUM, Y .
JOURNAL OF BIOMECHANICS, 1979, 12 (01) :83-93
[2]   Wearable sensors/systems and their impact on biomedical engineering - An overview from the guest editor [J].
Bonato, P .
IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 2003, 22 (03) :18-20
[3]   Surface-marker cluster design criteria for 3-D bone movement reconstruction [J].
Cappozzo, A ;
Cappello, A ;
DellaCroce, U ;
Pensalfini, F .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1997, 44 (12) :1165-1174
[4]  
CAPPOZZO A, 1982, THESIS U STRATHCLYDE
[5]  
FURNEE H, 1997, 3 DIMENSIONAL ANAL H, P85
[6]  
*GAL MOT CONTR INC, 1997, DMC 1400 SER MAN
[7]   Is it feasible to reconstruct body segment 3-D position and orientation using accelerometric data? [J].
Giansanti, D ;
Macellari, V ;
Maccioni, G ;
Cappozzo, A .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2003, 50 (04) :476-483
[8]  
GIANSANTI D, 2001, P INT S BIOM
[9]  
GOLDSTEIN H, 1977, CLASSICAL MECH
[10]   Accelerometer and rate gyroscope measurement of kinematics: an inexpensive alternative to optical motion analysis systems [J].
Mayagoitia, RE ;
Nene, AV ;
Veltink, PH .
JOURNAL OF BIOMECHANICS, 2002, 35 (04) :537-542