Application of a test package in an intercomparison of the photon dose calculation performance of treatment planning systems used in a clinical setting

被引:50
作者
Venselaar, J
Welleweerd, H
机构
[1] Dr B Verbeeten Inst, Dept Radiotherapy, NL-5000 LA Tilburg, Netherlands
[2] Univ Med Ctr, Dept Radiotherapy, NL-3508 GA Utrecht, Netherlands
关键词
treatment planning systems; quality assurance; dose calculation;
D O I
10.1016/S0167-8140(01)00304-8
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: Testing the performance of treatment planning systems by using the AAPM Task Group 23 test package is a useful approach, but has its limitations. To be able to include technical developments, such as the asymmetric collimator, it was decided to remeasure the AAPM data set on more modern radiotherapy equipment, to extend the test geometries, and to evaluate the use of the new package. Materials and methods: A coherent set of beam data of 6, 10 and 18 MV photon beams was measured on two modern linear accelerators. These data served as input data in seven commercially available treatment planning systems, which were clinically in use in different radiotherapy departments. Next, a test package was measured which included a missing tissue geometry and fields with asymmetrical collimator setting, with and without a wedge. Results: The absolute dose prediction from the different treatment planning systems in which the measured beam data were entered, was compared for all test points with the results of direct measurements. The criteria of acceptability were exceeded by some systems in cases of irregular field geometry and missing tissue geometry. The majority of the systems had difficulties with accurate dose calculation for asymmetrically wedged fields. Conclusions: The application of the new test package did not introduce insuperable difficulties and was highly appreciated by the participating centres. Most systems performed reasonably well for the majority of the beam geometries, with the exception of asymmetrically wedged beams. The extended test package is available for other users or user group, for the purpose of commissioning new treatment planning systems, or new releases of existing systems. (C) 2001 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:203 / 213
页数:11
相关论文
共 32 条
[1]   Application of AAPM Radiation Therapy Committee Task Group 23 test package for comparison of two treatment planning systems for photon external beam radiotherapy [J].
Alam, R ;
Ibbott, GS ;
Pourang, R ;
Nath, R .
MEDICAL PHYSICS, 1997, 24 (12) :2043-2054
[2]  
*AM ASS PHYS MED, 1995, 55 AAPM AM I PHYS
[3]  
Brahme A., 1988, ACTA ONCOL S, V1, P1
[4]   A method to check the accuracy of dose computation using quality index: Application to scatter contribution in high energy photon beams [J].
Caneva, S ;
Rosenwald, JC ;
Zefkili, S .
MEDICAL PHYSICS, 2000, 27 (05) :1018-1024
[5]   USER REQUIREMENTS ON CT-BASED COMPUTED DOSE PLANNING SYSTEMS IN RADIATION-THERAPY - PRESENTATION OF CHECK LISTS [J].
DAHLIN, H ;
LAMM, IL ;
LANDBERG, T ;
LEVERNES, S ;
ULSO, N .
COMPUTER PROGRAMS IN BIOMEDICINE, 1983, 16 (03) :131-137
[6]  
DAHLIN H, 1983, ACTA RADIOL ONCOL, V22, P396
[8]   Accuracy of two- and three-dimensional photon dose calculation for tangential irradiation of the breast [J].
Davis, JB ;
Pfafflin, A ;
Cozzi, AF .
RADIOTHERAPY AND ONCOLOGY, 1997, 42 (03) :245-248
[9]   Dosimetric evaluation of a commercial 3-D treatment planning system using Report 55 by AAPM Task Group 23 [J].
Declich, F ;
Fumasoni, K ;
Mangili, P ;
Cattaneo, GM ;
Iori, M .
RADIOTHERAPY AND ONCOLOGY, 1999, 52 (01) :69-77
[10]   Anthropomorphic phantom measurements for the validation of a treatment planning system [J].
Dunscombe, P ;
McGhee, P ;
Lederer, E .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (03) :399-411