Effect of undersizing on the long-term stability of the Exeter hip stem: A comparative in vitro study

被引:16
作者
Cristofolini, Luca [1 ,2 ]
Erani, Paolo [1 ]
Bialoblocka-Juszczyk, Ewa [1 ,2 ]
Ohashi, Hirotsugu [3 ]
Iida, Satoshi [4 ]
Minato, Izumi [5 ]
Viceconti, Marco [1 ]
机构
[1] Ist Ortoped Rizzoli, Lab Tecnol Med, I-40136 Bologna, Italy
[2] Univ Bologna, DIEM, I-40136 Bologna, Italy
[3] Saiseikai Nakatsu Hosp, Dept Orthopaed Surg, Kita Ku, Osaka 5300012, Japan
[4] Matsudo City Hosp, Dept Orthopaed Surg, Chiba 2710064, Japan
[5] Niigata Rinko Hosp, Dept Orthopaed Surg, Niigata 9500051, Japan
关键词
Polished cemented hip prosthesis; Stem undersizing; Long-term implant-bone stability; Fatigue failure; Inducible micromotion; Permanent migration; In vitro testing; CEMENTED FEMORAL COMPONENT; PRECLINICAL ASSESSMENT; FINITE-ELEMENT; INITIAL STABILITY; PATIENT ACTIVITY; SURFACE FINISH; MANTLE; ARTHROPLASTY; BONE; PROSTHESES;
D O I
10.1016/j.clinbiomech.2010.07.003
中图分类号
R318 [生物医学工程];
学科分类号
100103 [病原生物学];
摘要
Background: Even for clinically successful hip stems such as the Exeter-V40 occasional failures are reported. It has been reported that sub-optimal pre-operative planning, leading to implant undersizing and/or thin cement mantle, can explain such failures. The scope of this study was to investigate whether stem undersizing and a thin cement mantle are sufficient to cause implant loosening. Methods: A comparative in vitro study was designed to compare hip implants prepared with optimal and smaller than optimal stem size. Exeter-V40, a highly polished cemented hip stem, was used in both cases. Tests were carried out simulating 24 years of activity of active hip patients. A multifaceted approach was taken: inducible and permanent micromotions were recorded throughout the test; cement micro-cracks were quantified using dye penetrants and statistically analyzed. Findings: The implants with an optimal stem size withstood the entire mechanical test, with low and stable inducible micromotions and permanent migrations during the test, and with moderate fatigue damage in the cement mantle after test completion. Conversely, the undersized specimens showed large and increasing micromotions. and failed after few loading cycles, because of macroscopic cracks in the proximal part of the cement mantle. While results for the optimal stem size are typical for stable hip stems, those for the undersize stem indicate a critical scenario. Interpretation: These results confirm that even a clinically successful hip prosthesis such as the Exeter-V40 is prone to early loosening if a stem smaller than the optimal size is implanted. (C) 2010 Elsevier Ltd. All rights reserved.
引用
收藏
页码:899 / 908
页数:10
相关论文
共 51 条
[1]
Early failure of modern cemented stems [J].
Barrack, RL .
JOURNAL OF ARTHROPLASTY, 2000, 15 (08) :1036-1050
[2]
Hip contact forces and gait patterns from routine activities [J].
Bergmann, G ;
Deuretzbacher, G ;
Heller, M ;
Graichen, F ;
Rohlmann, A ;
Strauss, J ;
Duda, GN .
JOURNAL OF BIOMECHANICS, 2001, 34 (07) :859-871
[3]
Effect of long-term physiological activity on the long-term stem stability of cemented hip arthroplasty: in vitro comparison of three commercial bone cements [J].
Bialoblocka-Juszczyk, E. ;
Cristofolini, L. ;
Erani, P. ;
Viceconti, M. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2010, 224 (H1) :53-65
[4]
BREUSCH SJ, 2000, DTSCH ARZTEBLATT, V97, P2030
[5]
Preclinical testing of femoral hip components: An experimental investigation with four prostheses [J].
Britton, JR ;
Prendergast, PJ .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2005, 127 (05) :872-880
[6]
Preclinical assessment of the long-term endurance of cemented hip stems. Part 1: effect of daily activities a comparison of two load histories [J].
Cristofolini, L. ;
Teutonico, A. Saponara ;
Savigni, P. ;
Erani, P. ;
Viceconti, M. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2007, 221 (H6) :569-584
[7]
Preclinical assessment of the long-term endurance of cemented hip stems.: Part 2:: in-vitro and ex-vivo fatigue damage of the cement mantle [J].
Cristofolini, L. ;
Erani, P. ;
Savigni, P. ;
Bordini, B. ;
Viceconti, M. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2007, 221 (H6) :585-599
[8]
A critical analysis of stress shielding evaluation of hip prostheses [J].
Cristofolini, L .
CRITICAL REVIEWS IN BIOMEDICAL ENGINEERING, 1997, 25 (4-5) :409-483
[9]
Mechanical validation of whole bone composite femur models [J].
Cristofolini, L ;
Viceconti, M ;
Cappello, A ;
Toni, A .
JOURNAL OF BIOMECHANICS, 1996, 29 (04) :525-535
[10]
Comparative in vitro study on the long term performance of cemented hip stems: validation of a protocol to discriminate between "good" and "bad" designs [J].
Cristofolini, L ;
Teutonico, AS ;
Monti, L ;
Cappello, A ;
Toni, A .
JOURNAL OF BIOMECHANICS, 2003, 36 (11) :1603-1615