Benchmarking beam alignment for a clinical helical tomotherapy device

被引:48
作者
Balog, J [1 ]
Mackie, TR
Pearson, D
Hui, S
Paliwal, B
Jeraj, R
机构
[1] TomoTherapy Inc, Madison, WI 53717 USA
[2] Univ Wisconsin, Dept Med Phys, Madison, WI 53706 USA
关键词
tomotherapy; commissioning dosimetry; beam alignment; multileaf collimator;
D O I
10.1118/1.1576395
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A clinical helical tomotherapy treatment machine has been installed at the University of Wisconsin Comprehensive Cancer Center. Beam alignment has been finalized and accepted by UW staff. Helical tomotherapy will soon be clinically available to other sites. Clinical physicists who expect to work with this machine will need to be familiar with its unique dosimetric characteristics, and those related to the geometrical beam configuration and its verification are described here. A series of alignment tests and the results are presented. Helical tomotherapy utilizes an array of post-patient xenon-filled megavoltage radiation detectors. These detectors have proved capable of performing some alignment verification tests. That is particularly advantageous because those tests can then be automated and easily performed on an ongoing basis. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:1118 / 1127
页数:10
相关论文
共 14 条
[11]   Tomotherapy [J].
Mackie, TR ;
Balog, J ;
Ruchala, K ;
Shepard, D ;
Aldridge, S ;
Fitchard, E ;
Reckwerdt, P ;
Olivera, G ;
McNutt, T ;
Mehta, M .
SEMINARS IN RADIATION ONCOLOGY, 1999, 9 (01) :108-117
[12]   TOMOTHERAPY - A NEW CONCEPT FOR THE DELIVERY OF DYNAMIC CONFORMAL RADIOTHERAPY [J].
MACKIE, TR ;
HOLMES, T ;
SWERDLOFF, S ;
RECKWERDT, P ;
DEASY, JO ;
YANG, J ;
PALIWAL, B ;
KINSELLA, T .
MEDICAL PHYSICS, 1993, 20 (06) :1709-1719
[13]  
Saw C B, 2001, Med Dosim, V26, P55, DOI 10.1016/S0958-3947(00)00056-X
[14]   Investigation of tomotherapy beam delivery [J].
Yang, JN ;
Mackie, TR ;
Reckwerdt, P ;
Deasy, JO ;
Thomadsen, BR .
MEDICAL PHYSICS, 1997, 24 (03) :425-436