Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energies

被引:142
作者
Chiu-Tsao, ST [1 ]
Ho, Y
Shankar, R
Wang, L
Harrison, LB
机构
[1] Beth Israel Deaconess Med Ctr, Dept Radiat Oncol, New York, NY 10003 USA
[2] St Lukes Roosevelt Hosp, New York, NY 10003 USA
关键词
radiochromic film; film sensitivity; absorption spectrum; energy response; light source;
D O I
10.1118/1.2065467
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
The purpose of this paper is to evaluate the energy dependence of the response of two new high sensitivity models of radiochromic films EBT and XR-QA. We determined the dose response curves of these films for four different radiation sources, namely, 6 MV photon beams (6 MVX), Ir-192, I-125, and Pd-103. The first type (EBT) is designed for intensity modulated radiation therapy (IMRT) dosimetry, and the second type (XR-QA) is designed for kilovoltage dosimetry. All films were scanned using red (665 nm) and green (520 nm) light sources in a charge-coupled device-based densitometer. The dose response curves [net optical density (NOD) versus dose] were plotted and compared for different radiation energies and light sources. Contrary to the early GAFCHRO-MIC film types (such as models XR, HS, MD55-2, and HD810), the net optical densities of both EBT and XR-QA were higher with a green (520 nm) than those with a red (665 nm) light source due to the different absorption spectrum of the new radiochromic emulsion. Both film types yield measurable optical densities for doses below 2 Gy. EBT film response is nearly independent of radiation energy, within the uncertainty of measurement. The NOD values of EBT film at 1 and 2 Gy are 0.13 and 0.25 for green, and 0.1 and 0.17 for red, respectively. In contrast, the XR-QA film sensitivity varies with radiation energy. The doses required to produce NOD of 0.5 are 6.9, 5.4, 0.7, and 0.9 Gy with green light and 19, 13, 1.7, and 1.5 Gy with red light, for 6 MVX, Ir-192, I-125, and Pd-103, respectively. EBT film was found to have minimal photon energy dependence of response for the energies tested and is suitable for dosimetry of radiation with a wide energy spectrum, including primary and scattered radiation. XR-QA film is promising for kilovoltage sources with a narrow energy spectra. The new high sensitivity radiochromic films are promising tools in radiation dosimetry. (c) 2005 American Association of Physicists in Medicine.
引用
收藏
页码:3350 / 3354
页数:5
相关论文
共 44 条
[1]
Dependence of radiochromic film optical density post-exposure kinetics on dose and dose fractionation [J].
Ali, I ;
Costescu, C ;
Vicic, M ;
Dempsey, JF ;
Williamson, JF .
MEDICAL PHYSICS, 2003, 30 (08) :1958-1967
[2]
A wide dynamic range, high-spatial-resolution scanning system for radiochromic dye films [J].
Aydarous, AS ;
Darley, PJ ;
Charles, MW .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (05) :1379-1389
[3]
Baker J, 2004, MED PHYS, V31, P1726
[4]
Measurements and Monte Carlo calculations to determine the absolute detector response of radiochromic film for brachytherapy dosimetry [J].
Bohm, TD ;
Pearson, DW ;
Das, RK .
MEDICAL PHYSICS, 2001, 28 (02) :142-146
[5]
Radiochromic film for medical radiation dosimetry [J].
Butson, MJ ;
Yu, PKN ;
Cheung, T ;
Metcalfe, P .
MATERIALS SCIENCE & ENGINEERING R-REPORTS, 2003, 41 (3-5) :61-120
[6]
High sensitivity radiochromic film dose comparisons [J].
Butson, MJ ;
Yu, PKN ;
Cheung, T ;
Metcalfe, P .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (22) :N291-N295
[7]
Experimental energy response verification of XR type T radiochromic film [J].
Cheung, T ;
Butson, MJ ;
Yu, PKN .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (21) :N371-N376
[8]
Evaluation of a fluorescent light densitometer for radiochromic film analysis [J].
Cheung, T ;
Butson, MJ ;
Yu, PKN .
RADIATION MEASUREMENTS, 2002, 35 (01) :13-16
[9]
Chiu-Tsao S, 2004, MED PHYS, V31, P1727
[10]
Dose response characteristics of new models of GAFCHROMIC films: Dependence on densitometer light source and radiation energy [J].
Chiu-Tsao, ST ;
Duckworth, T ;
Zhang, CF ;
Patel, NS ;
Hsiung, CY ;
Lin, W ;
Shih, JA ;
Harrison, LB .
MEDICAL PHYSICS, 2004, 31 (09) :2501-2508