Reversible electrochemical detection of nonelectroactive polyions

被引:149
作者
Shvarev, A [1 ]
Bakker, E [1 ]
机构
[1] Auburn Univ, Dept Chem, Auburn, AL 36849 USA
关键词
D O I
10.1021/ja037167n
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Selective extraction principles for the recognition of nonelectroactive polyions such as heparin and protamine exist, but the high ionic valency renders the extraction process irreversible. A response principle for the reversible detection of such polyions is proposed here. The extraction of the polyionic analyte to the membrane and its subsequent back-extraction is now controlled electrochemically. The principle is established with a protamine electrode, and excellent stability and reproducibility are demonstrated. This method has important implications for the design of chemical recognition principles for polyionic analytes. Copyright © 2003 American Chemical Society.
引用
收藏
页码:11192 / 11193
页数:2
相关论文
共 12 条
[1]   Direct potentiometric information on total ionic concentrations [J].
Ceresa, A ;
Pretsch, E ;
Bakker, E .
ANALYTICAL CHEMISTRY, 2000, 72 (09) :2050-2054
[2]   EXTRACTION THERMODYNAMICS OF POLYANIONS INTO PLASTICIZED POLYMER MEMBRANES DOPED WITH LIPOPHILIC ION-EXCHANGERS - A POTENTIOMETRIC STUDY [J].
FU, B ;
BAKKER, E ;
YANG, VC ;
MEYERHOFF, ME .
MACROMOLECULES, 1995, 28 (17) :5834-5840
[3]   RESPONSE MECHANISM OF POLYMER MEMBRANE-BASED POTENTIOMETRIC POLYION SENSORS [J].
FU, B ;
BAKKER, E ;
YUN, JH ;
YANG, VC ;
MEYERHOFF, ME .
ANALYTICAL CHEMISTRY, 1994, 66 (14) :2250-2259
[4]   HEPARIN-RESPONSIVE ELECTROCHEMICAL SENSOR - A PRELIMINARY-STUDY [J].
MA, SC ;
YANG, VC ;
MEYERHOFF, ME .
ANALYTICAL CHEMISTRY, 1992, 64 (06) :694-697
[5]   Renewable pH cross-sensitive potentiometric heparin sensors with incorporated electrically charged H+ ionophores [J].
Mathison, S ;
Bakker, E .
ANALYTICAL CHEMISTRY, 1999, 71 (20) :4614-4621
[6]   Polyion-sensitive membrane electrodes for biomedical analysis [J].
Meyerhoff, ME ;
Fu, B ;
Bakker, E ;
Yun, JH ;
Yang, VC .
ANALYTICAL CHEMISTRY, 1996, 68 (05) :A168-A175
[7]   Polyion sensors as liquid junction-free reference electrodes [J].
Mi, YM ;
Mathison, S ;
Bakker, E .
ELECTROCHEMICAL AND SOLID STATE LETTERS, 1999, 2 (04) :198-200
[8]   Improved detection limits and sensitivities of potentiometric titrations [J].
Peper, S ;
Ceresa, A ;
Bakker, E ;
Pretsch, E .
ANALYTICAL CHEMISTRY, 2001, 73 (15) :3768-3775
[9]  
Ramamurthy N, 1998, CLIN CHEM, V44, P606
[10]   Pulsed galvanostatic control of ionophore-based polymeric ion sensors [J].
Shvarev, A ;
Bakker, E .
ANALYTICAL CHEMISTRY, 2003, 75 (17) :4541-4550