The accuracy of dose localization for an image-guided frameless radiosurgery system

被引:108
作者
Murphy, MJ
Cox, RS
机构
[1] Department of Radiation Oncology, Stanford University, School of Medicine, Stanford
关键词
radiosurgery; radiotherapy; robotics;
D O I
10.1118/1.597771
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A compact X-band accelerator mounted on a robotic arm is under development for frameless stereotaxic radiosurgery. The therapy beam is aimed at the lesion by an imaging system comprised of two diagnostic x-ray cameras that view the patient during treatment. Patient position and motion are measured by the cameras and communicated in real time to the robotic arm for beam targeting and patient motion tracking. The tests reported here measured the pointing accuracy of the therapy beam and the present targeting and tracking capability of the imaging system. The results show that the system achieves the same level of targeting precision as conventional frame-based radiosurgery. (C) 1996 American Association of Physicists in Medicine.
引用
收藏
页码:2043 / 2049
页数:7
相关论文
共 7 条
  • [1] VARIABLES AFFECTING THE ACCURACY OF STEREOTAXIC LOCALIZATION USING COMPUTERIZED-TOMOGRAPHY
    BUCHOLZ, RD
    HO, HW
    RUBIN, JP
    [J]. JOURNAL OF NEUROSURGERY, 1993, 79 (05) : 667 - 673
  • [2] THE UNIVERSITY-OF-FLORIDA RADIOSURGERY SYSTEM
    FRIEDMAN, WA
    BOVA, FJ
    [J]. SURGICAL NEUROLOGY, 1989, 32 (05): : 334 - 342
  • [3] THE APPLICATION ACCURACY OF STEREOTAXIC FRAMES
    MACIUNAS, RJ
    GALLOWAY, RL
    LATIMER, JW
    [J]. NEUROSURGERY, 1994, 35 (04) : 682 - 694
  • [4] THE USE OF A RADIOCHROMIC DETECTOR FOR THE DETERMINATION OF STEREOTAXIC RADIOSURGERY DOSE CHARACTERISTICS
    MCLAUGHLIN, WL
    SOARES, CG
    SAYEG, JA
    MCCULLOUGH, EC
    KLINE, RW
    WU, A
    MAITZ, AH
    [J]. MEDICAL PHYSICS, 1994, 21 (03) : 379 - 388
  • [5] MURPHY MJ, 1996, MED PHYS, V23, P1052
  • [6] THE USE OF RADIOCHROMIC FILM IN TREATMENT VERIFICATION OF DYNAMIC STEREOTAXIC RADIOSURGERY
    RAMANI, R
    LIGHTSTONE, AW
    MASON, DLD
    OBRIEN, PF
    [J]. MEDICAL PHYSICS, 1994, 21 (03) : 389 - 392
  • [7] LINEAR-ACCELERATOR AS A NEUROSURGICAL TOOL FOR STEREOTACTIC RADIOSURGERY
    WINSTON, KR
    LUTZ, W
    [J]. NEUROSURGERY, 1988, 22 (03) : 454 - 464