Dose reduction in spiral CT angiography of thoracic outlet syndrome by anatomically adapted tube current modulation

被引:33
作者
Mastora, I
Remy-Jardin, M
Suess, C
Scherf, C
Guillot, JP
Remy, J
机构
[1] Univ Ctr Hosp Calmette, Dept Radiol, F-59037 Lille, France
[2] Med Res Grp Equipe, F-59037 Lille, France
[3] Siemens Med Syst Inc, D-91301 Forcheim, Germany
关键词
CT dose; dose reduction; image quality; spiral CT angiography;
D O I
10.1007/s003300000752
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was to evaluate dose reduction in spiral CT angiography of the thoracic outlet by on-line tube-current control. Prospectively, 114 patients undergoing spiral CT angiography of the subclavian artery for thoracic outlet arterial syndromes were evaluated with and without tube-current modulation at the same session (scanning parameters for the two successive angiograms, one in the neutral position and one after the postural maneuver): 140 kV; 206 mA: scan time 0.75 s; collimation 3 mm; pitch = 1). The dose reduction system was applied in the neutral position in the first 92 consecutive patients and after postural maneuver in the remaining 22 consecutive patients. Dose reduction and image quality were analyzed in the overall study group (group 1; n = 114). The influence of the arm position was assessed in 44 of the 114 patients (group 2), matched by the transverse diameter of the upper thorax. The mean dose reduction was 33 % in group 1 (range 22-40 %) and 34 % in group 2 (range 26-40 %). In group 2 the only difference in image quality was a significantly higher frequency of graininess on low-dose scans compared with reference scans whatever the patient's arm position, graded as minimal in 38 of the 44 patients (86 %). When the low-dose technique was applied after postural maneuver in group 2: (a) the mean dose reduction was significantly higher (35 vs 32 % in the neutral position; p = 0.006); (b) graininess was less frequent (82 vs 91% in the neutral position); and (c) the percentage of graininess graded as minimal was significantly higher (83 vs 70 % in the neutral position; p = 0.2027). On-line tube-current modulation enables dose reduction on high-quality, diagnostic spiral CT angiograms of the thoracic outlet and should be applied during data acquisition in the neutral position and after postural maneuver for optimal use.
引用
收藏
页码:590 / 596
页数:7
相关论文
共 13 条
[1]  
*ALARA NAT COUNC R, 1990, 107 NCRP ALARA
[2]   HELICAL CT - PRINCIPLES AND TECHNICAL CONSIDERATIONS [J].
BRINK, JA ;
HEIKEN, JP ;
WANG, G ;
MCENERY, KW ;
SCHLUETER, FJ ;
VANNIER, MW .
RADIOGRAPHICS, 1994, 14 (04) :887-893
[3]   Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions [J].
Greess, H ;
Wolf, H ;
Baum, U ;
Lell, M ;
Pirkl, M ;
Kalender, W ;
Bautz, WA .
EUROPEAN RADIOLOGY, 2000, 10 (02) :391-394
[4]   Reducing data size and radiation dose for CT colonography [J].
Hara, AK ;
Johnson, CD ;
Reed, JE ;
Ahlquist, DA ;
Nelson, H ;
Ehman, RL ;
Harmsen, WS .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1997, 168 (05) :1181-1184
[5]   Dose reduction in CT by on-line tube current control: principles and validation on phantoms and cadavers [J].
Kalender, WA ;
Wolf, H ;
Suess, C ;
Gies, M ;
Greess, H ;
Bautz, WA .
EUROPEAN RADIOLOGY, 1999, 9 (02) :323-328
[6]   Effective doses to members of the public from the diagnostic application of ionizing radiation in Germany [J].
Kaul, A ;
Bauer, B ;
Bernhardt, J ;
Nosske, D ;
Veit, R .
EUROPEAN RADIOLOGY, 1997, 7 (07) :1127-1132
[7]  
Luoma A, 1991, Neurosurg Clin N Am, V2, P187
[8]   GT OF THE CHEST - MINIMAL TUBE CURRENT REQUIRED FOR GOOD IMAGE QUALITY WITH THE LEAST RADIATION-DOSE [J].
MAYO, JR ;
HARTMAN, TE ;
LEE, KS ;
PRIMACK, SL ;
VEDAL, S ;
MULLER, NL .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 164 (03) :603-607
[9]   LOW-DOSE CT OF THE LUNGS - PRELIMINARY-OBSERVATIONS [J].
NAIDICH, DP ;
MARSHALL, CH ;
GRIBBIN, C ;
ARAMS, RS ;
MCCAULEY, DI .
RADIOLOGY, 1990, 175 (03) :729-731
[10]   EVALUATION OF SECTION SENSITIVITY PROFILES AND IMAGE NOISE IN SPIRAL CT [J].
POLACIN, A ;
KALENDER, WA ;
MARCHAL, G .
RADIOLOGY, 1992, 185 (01) :29-35