Partial volume effect correction in PET using regularized iterative deconvolution with variance control based on local topology

被引:55
作者
Kirov, A. S. [1 ]
Piao, J. Z. [2 ]
Schmidtlein, C. R. [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, New York, NY 11021 USA
[2] Cleveland Clin, Cleveland, OH 44195 USA
关键词
D O I
10.1088/0031-9155/53/10/009
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Correcting positron emission tomography (PET) images for the partial volume effect (PVE) due to the limited resolution of PET has been a long-standing challenge. Various approaches including incorporation of the system response function in the reconstruction have been previously tested. We present a post-reconstruction PVE correction based on iterative deconvolution using a 3D maximum likelihood expectation-maximization (MLEM) algorithm. To achieve convergence we used a one step late (OSL) regularization procedure based on the assumption of local monotonic behavior of the PET signal following Alenius et al. This technique was further modified to selectively control variance depending on the local topology of the PET image. No prior 'anatomic' information is needed in this approach. An estimate of the noise properties of the image is used instead. The procedure was tested for symmetric and isotropic deconvolution functions with Gaussian shape and full width at half-maximum (FWHM) ranging from 6.31 mm to infinity. The method was applied to simulated and experimental scans of the NEMA NU 2 image quality phantom with the GE Discovery LS PET/CT scanner. The phantom contained uniform activity spheres with diameters ranging from 1 cm to 3.7 cm within uniform background. The optimal sphere activity to variance ratiowas obtained when the deconvolution function was replaced by a step function few voxels wide. In this case, the deconvolution method converged in similar to 3-5 iterations for most points on both the simulated and experimental images. For the 1 cm diameter sphere, the contrast recovery improved from 12% to 36% in the simulated and from 21% to 55% in the experimental data. Recovery coefficients between 80% and 120% were obtained for all larger spheres, except for the 13 mm diameter sphere in the simulated scan (68%). No increase in variance was observed except for a few voxels neighboring strong activity gradients and inside the largest spheres. Testing the method for patient images increased the visibility of small lesions in non-uniform background and preserved the overall image quality. Regularized iterative deconvolution with variance control based on the local properties of the PET image and on estimated image noise is a promising approach for partial volume effect corrections in PET.
引用
收藏
页码:2577 / 2591
页数:15
相关论文
共 29 条
[1]   Bayesian image reconstruction for emission tomography based on median root prior [J].
Sakari Alenius ;
Ulla Ruotsalainen .
European Journal of Nuclear Medicine, 1997, 24 (3) :258-265
[2]  
ALENIUS S, 1999, THESIS TAMPERE U TEC
[3]   Analytical reconstruction of deconvolved Fourier rebinned PET sinograms [J].
Alessio, A ;
Sauer, K ;
Kinahan, P .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (01) :77-93
[4]   Improved quantitation for PET/CT image reconstruction with system modeling and anatomical priors [J].
Alessio, Adam M. ;
Kinahan, Paul E. .
MEDICAL PHYSICS, 2006, 33 (11) :4095-4103
[5]  
[Anonymous], NEMA STAND PUBL NU
[6]   A multiresolution image based approach for correction of partial volume effects in emission tomography [J].
Boussion, N ;
Hatt, M ;
Lamare, F ;
Bizais, Y ;
Turzo, A ;
Rest, CCL ;
Visvikis, D .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (07) :1857-1876
[7]  
BOUSSION N, 2007, IEEE NUCL SCI S MED
[9]   APPLICATION OF MATHEMATICAL REMOVAL OF POSITRON RANGE BLURRING IN POSITRON EMISSION TOMOGRAPHY [J].
HABER, SF ;
DERENZO, SE ;
UBER, D .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (03) :1293-1299
[10]   QUANTITATION IN POSITRON EMISSION COMPUTED-TOMOGRAPHY .6. EFFECT OF NONUNIFORM RESOLUTION [J].
HOFFMAN, EJ ;
HUANG, SC ;
PLUMMER, D ;
PHELPS, ME .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1982, 6 (05) :987-999