Computer navigation versus standard instrumentation for TKA - A single-surgeon experience

被引:118
作者
Bolognesi, M
Hofmann, A
机构
[1] Univ Utah, Med Ctr, Dept Orthoped Surg, Salt Lake City, UT 84132 USA
[2] Duke Univ, Med Ctr, Div Orthopaed Surg, Durham, NC 27710 USA
关键词
D O I
10.1097/01.blo.0000186561.70566.95
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Component alignment errors in total knee arthroplasty greater than 3 degrees can be associated with poorer outcomes. This retrospective study seeks to determine if computer navigation can improve accuracy of component alignment in comparable patient populations. The efficiency and safety of the navigated technique is also evaluated. Fifty total knee arthroplasties done using an imageless navigation system and 50 cases using standard instrumentation were compared. The same surgeon used a single system (Zimmer-Natural Knee (TM)) in all cases. Long-standing radiographs collected at 6-week followup were measured for component orientation. When the navigation system was used 98% (49 of 50 cases) of all femoral components and 100% (50 of 50 cases) of all tibial components were placed within +/- 3 degrees of the radiographic goal position. There was a decrease in the standard instrumentation group to 90% (45 of 50 cases) and 92% (46 of 50 cases) within +/- 3 degrees, respectively. There was a difference in the standard deviations observed for the navigated cases and the conventional cases when femoral and tibial component position was considered. Average tourniquet time was 68 minutes in the navigated group and 57 minutes in the conventional group. There were no technique specific complications associated with the navigation system. This system affords the surgeon the potential to reduce outliers with regard to component position without an increase in complications. Tourniquet times were increased with the use of the computer. Level of Evidence: Therapeutic study, Level III-1 (retrospective comparative study). See the Guidelines for authors for a complete description of levels of evidence.
引用
收藏
页码:162 / 169
页数:8
相关论文
共 47 条
[1]   Computed tomography-based navigation for hip, knee, and spine surgery [J].
Amiot, LP ;
Poulin, F .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (421) :77-86
[2]   Radiological results of image-based and non-image-based computer-assisted total knee arthroplasty [J].
Bäthis, H ;
Perlick, L ;
Tingart, M ;
Lüring, C ;
Perlick, C ;
Grifka, J .
INTERNATIONAL ORTHOPAEDICS, 2004, 28 (02) :87-90
[3]  
Bäthis H, 2004, ORTHOPEDICS, V27, P476
[4]   Tibial component failure mechanisms in total knee arthroplasty [J].
Berend, ME ;
Ritter, MA ;
Meding, JB ;
Faris, PM ;
Keating, EM ;
Redelman, R ;
Faris, GW ;
Davis, KE .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (428) :26-34
[5]  
CAILLOUETTE JT, 1990, CLIN ORTHOP RELAT R, P198
[6]  
CATES HE, 1993, CLIN ORTHOP RELAT R, P32
[7]  
Clemens Ulrich, 2003, Surg Technol Int, V11, P265
[8]  
Delp SL, 1998, CLIN ORTHOP RELAT R, P49
[9]  
Dennis D A, 1993, J Arthroplasty, V8, P43, DOI 10.1016/S0883-5403(06)80106-3
[10]  
DORR LD, 1986, CLIN ORTHOP RELAT R, P5