The use of an inexpensive red acetate filter to improve the sensitivity of GAFChromic dosimetry

被引:31
作者
Odero, DO [1 ]
Gluckman, GR [1 ]
Welsh, K [1 ]
Wlodarczyk, RA [1 ]
Reinstein, LE [1 ]
机构
[1] SUNY Stony Brook, Dept Radiat Oncol, Stony Brook, NY 11794 USA
关键词
film digitizer; film digitization; film scanners; film densitometry; film dosimetry; film filters;
D O I
10.1118/1.1382605
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The sensitivity of GAFChromic dosimetry using a conventional broad band light source densitometer has significantly been improved twofold using an inexpensive red acetate filter overlay during the densitometric measurements. This thin sheet of red acetate enhances the dosimetric analysis of radiochromic blue image distributions recorded on GAFChromic films. The combination provides higher sensitivity in the optical density measurements than the more expensive He-Ne laser-scanning densitometers. (C) 2001 American Association of Physicists in Medicine.
引用
收藏
页码:1446 / 1448
页数:3
相关论文
共 10 条
[1]   Ultraviolet radiation dosimetry with radiochromic film [J].
Butson, MJ ;
Cheung, T ;
Yu, PKN ;
Abbati, D ;
Greenoak, GE .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (07) :1863-1868
[2]   Novel radiochromic films for clinical dosimetry [J].
McLaughlin, WL ;
Puhl, JM ;
AlSheikhly, M ;
Christou, CA ;
Miller, A ;
Kovacs, A ;
Wojnarovits, L ;
Lewis, DF .
RADIATION PROTECTION DOSIMETRY, 1996, 66 (1-4) :263-268
[3]  
McLaughlin WL, 1996, ACS SYM SER, V620, P152
[4]   SENSITOMETRY OF THE RESPONSE OF A NEW RADIOCHROMIC FILM DOSIMETER TO GAMMA-RADIATION AND ELECTRON-BEAMS [J].
MCLAUGHLIN, WL ;
YUNDONG, C ;
SOARES, CG ;
MILLER, A ;
VANDYK, G ;
LEWIS, DF .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 302 (01) :165-176
[5]   TESTS FOR EVALUATING LASER FILM DIGITIZERS [J].
MEEDER, RJJ ;
JAFFRAY, DA ;
MUNRO, P .
MEDICAL PHYSICS, 1995, 22 (05) :635-642
[6]   Characteristics of a commercially available film digitizer and their significance for film dosimetry [J].
Mersseman, B ;
De Wagter, C .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (06) :1803-1812
[7]   Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55 [J].
Niroomand-Rad, A ;
Blackwell, CR ;
Coursey, BM ;
Gall, KP ;
Galvin, JM ;
McLaughlin, WL ;
Meigooni, AS ;
Nath, R ;
Rodgers, JE ;
Soares, CG .
MEDICAL PHYSICS, 1998, 25 (11) :2093-2115
[8]   Predicting optical densitometer response as a function of light source characteristics for radiochromic film dosimetry [J].
Reinstein, LE ;
Gluckman, GR ;
Amols, HI .
MEDICAL PHYSICS, 1997, 24 (12) :1935-1942
[9]  
SEQUEIRARODRIGU.C, COMMUNICATION
[10]   Quantitative evaluation of radiochromic film response for two-dimensional dosimetry [J].
Zhu, YM ;
Kirov, AS ;
Mishra, V ;
Meigooni, AS ;
Williamson, JF .
MEDICAL PHYSICS, 1997, 24 (02) :223-231