Evaluation of the precision of magnetic resonance phase velocity mapping for blood flow measurements

被引:83
作者
Chatzimavroudis, GP
Oshinski, JN
Franch, RH
Walker, PG
Yoganathan, AP
Pettigrew, RI
机构
[1] Cleveland State Univ, Dept Chem Engn, Cleveland, OH 44115 USA
[2] Emory Univ, Sch Med, Dept Radiol, Frederik Philips Magnet Resonance Res Ctr, Atlanta, GA 30322 USA
[3] Emory Univ, Sch Med, Dept Med, Div Cardiol, Atlanta, GA 30322 USA
[4] Univ Leeds, Dept Mech Engn, Leeds LS2 9JT, W Yorkshire, England
[5] Georgia Inst Technol, Dept Biomed Engn, Atlanta, GA 30332 USA
关键词
accuracy; aortic regurgitation; flow quantification; magnetic resonance; magnetic resonance phase velocity mapping; precision;
D O I
10.1081/JCMR-100000142
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Evaluating the in vivo accuracy of magnetic resonance phase velocity. mapping (PVM) is not straightforward because of the absence of a validated clinical flow quantification technique. The aim of this study was to evaluate PVM by investigating its precision, both in vitro and in vivo, in a 1.5 Tesla scanner In the former case, steady and pulsatile flow experiments were conducted using an aortic model under a variety of flow conditions (steady : 0.1-5.5 L/min, pulsatile. 10-75 mL/cycle). In the latter case, PVM measurements were taken in the ascending aorta often subjects, seven of which hem aortic regurgitation. Each velocity measurement was taken twice, with the slice perpendicular to the long axis of the aorta. Comparison between, the measured and true flow rates and volumes confirmed the high accuracy of PVM in measuring pou in vitro (p > 0.85). The in vitro precision of PVM was found to be very high (steady: y = 1.00x + 0.02. r = 0.999: pulsatile: y = 0.98x + 0.72. r = 0.997: x: measurement #1, y : measurement #2) and this was confirmed by Bland-Airman analysis. Of great clinical significance was the high level of the in vivo precision (y = 1.01x - 0.04. r = 0.993), confirmed statistically (p = 1.00). In conclusion. PVM provides repeatable bloodflow measurements. The high in vitro accuracy and precision, combined with the high in vivo precision, are key factors for the establishment of PVM as the "gold-standard" to quantify bloodflow.
引用
收藏
页码:11 / 19
页数:9
相关论文
共 22 条
[1]   BLOOD-FLOW IN THE CAROTID ARTERIES - QUANTIFICATION BY USING PHASE-SENSITIVE MR IMAGING [J].
BENDEL, P ;
BUONOCORE, E ;
BOCKISCH, A ;
BESOZZI, MC .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1989, 152 (06) :1307-1310
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   BLOOD-FLOW MEASUREMENTS IN THE AORTA AND MAJOR ARTERIES WITH MR VELOCITY MAPPING [J].
BOGREN, HG ;
BUONOCORE, MH .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 1994, 4 (02) :119-130
[4]   MAGNETIC-RESONANCE-IMAGING COMPARED TO ECHOCARDIOGRAPHY TO ASSESS LEFT-VENTRICULAR MASS IN THE HYPERTENSIVE PATIENT [J].
BOTTINI, PB ;
CARR, AA ;
PRISANT, LM ;
FLICKINGER, FW ;
ALLISON, JD ;
GOTTDIENER, JS .
AMERICAN JOURNAL OF HYPERTENSION, 1995, 8 (03) :221-228
[5]   MEASUREMENT OF FLOW WITH NMR IMAGING USING A GRADIENT PULSE AND PHASE DIFFERENCE TECHNIQUE [J].
BRYANT, DJ ;
PAYNE, JA ;
FIRMIN, DN ;
LONGMORE, DB .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1984, 8 (04) :588-593
[6]   The importance of slice location on the accuracy of aortic regurgitation measurements with magnetic resonance phase velocity mapping [J].
Chatzimavroudis, GP ;
Walker, PG ;
Oshinski, JN ;
Franch, RH ;
Pettigrew, RI ;
Yoganathan, AP .
ANNALS OF BIOMEDICAL ENGINEERING, 1997, 25 (04) :644-652
[7]  
Chatzimavroudis GP, 1998, J HEART VALVE DIS, V7, P94
[8]   Slice location dependence of aortic regurgitation measurements with MR phase velocity mapping [J].
Chatzimavroudis, GP ;
Walker, PG ;
Oshinski, JN ;
Franch, RH ;
Pettigrew, RI ;
Yoganathan, AP .
MAGNETIC RESONANCE IN MEDICINE, 1997, 37 (04) :545-551
[9]   Adaptive correction of imaging plane position in segmented K-space cine cardiac MRI [J].
Chuang, ML ;
Chen, MH ;
Khasgiwala, VC ;
McConnell, MV ;
Edelman, RR ;
Manning, WJ .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1997, 7 (05) :811-814
[10]   SEVERITY OF AORTIC REGURGITATION - INTERSTUDY REPRODUCIBILITY OF MEASUREMENTS WITH VELOCITY-ENCODED CINE MR-IMAGING [J].
DULCE, MC ;
MOSTBECK, GH ;
OSULLIVAN, M ;
CHEITLIN, M ;
CAPUTO, GR ;
HIGGINS, CB .
RADIOLOGY, 1992, 185 (01) :235-240