Comparison of the absolute detector-based spectral radiance assignment with the current NIST-assigned spectral radiance of tungsten strip lamps

被引:8
作者
Yoon, HW [1 ]
Gibson, CE [1 ]
机构
[1] NIST, Opt Technol Div, Gaithersburg, MD 20899 USA
关键词
D O I
10.1088/0026-1394/37/5/18
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Using a high-temperature black body (HTBB) and filter radiometers calibrated for absolute spectral power responsivity, the spectral output of the black body, whose radiance temperature was determined using the filter radiometers, was used to assign spectral radiance to tungsten strip lamps with a prism-grating monochromator. The spectral radiance of the HTBB was also determined using signal ratios to a tungsten strip lamp calibrated using a scale derived from the radiometric temperature determination of a gold-freezing-point black body. The radiance temperatures found using the two methods were in agreement to 0.5 K near 2900 K, and from 260 nm to 1050 nm, the spectral radiance of the HTBB did not differ more than 0.5 % in radiance from a single-temperature Planck's law as determined using the tungsten strip lamp.
引用
收藏
页码:429 / 432
页数:4
相关论文
共 11 条
[1]  
[Anonymous], P TEMPEMKO 1999 7 IN
[2]   High-accuracy aperture-area measurement facilities at the National Institute of Standards and Technology [J].
Fowler, JB ;
Durvasula, RS ;
Parr, AC .
METROLOGIA, 1998, 35 (04) :497-500
[3]   International comparison of radiation-temperature measurements with filtered detectors over the temperature range 1380 K to 3100 K [J].
Harrison, NJ ;
Fox, NP ;
Sperfeld, P ;
Metzdorf, J ;
Khlevnoy, BB ;
Stolyarevskaya, RI ;
Khromchenko, VB ;
Mekhontsev, SN ;
Shapoval, VI ;
Zelener, MF ;
Sapritsky, VI .
METROLOGIA, 1998, 35 (04) :283-288
[4]   A METHOD OF REALIZING SPECTRAL IRRADIANCE BASED ON AN ABSOLUTE CRYOGENIC RADIOMETER [J].
JOHNSON, BC ;
CROMER, CL ;
SAUNDERS, RD ;
EPPELDAUER, G ;
FOWLER, J ;
SAPRITSKY, VI ;
DEZSI, G .
METROLOGIA, 1993, 30 (04) :309-315
[5]  
LARASON TC, 1998, NATL I STAND TECHNOL
[6]   SPECTRORADIOMETRIC DETERMINATION OF THE FREEZING TEMPERATURE OF GOLD [J].
MIELENZ, KD ;
SAUNDERS, RD ;
SHUMAKER, JB .
JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, 1990, 95 (01) :49-67
[7]   APPARATUS FUNCTION OF A PRISM-GRATING DOUBLE MONOCHROMATOR [J].
SAUNDERS, RD ;
SHUMAKER, JB .
APPLIED OPTICS, 1986, 25 (20) :3710-3714
[8]   Investigation of high-temperature black body BB3200 [J].
Sperfeld, P ;
Metzdorf, J ;
Harrison, NJ ;
Fox, NP ;
Khlevnoy, BB ;
Khromchenko, VB ;
Mekhontsev, SN ;
Shapoval, VI ;
Zelener, MF ;
Sapritsky, VI .
METROLOGIA, 1998, 35 (04) :419-422
[9]   Comparison of filter radiometer spectral responsivity with the NIST spectral-irradiance and illuminance scales [J].
Tsai, BK ;
Johnson, BC ;
Saunders, RD ;
Cromer, CL .
METROLOGIA, 1996, 32 (06) :473-477
[10]  
WALKER JH, 1987, NBS US SPEC PUBL