Chemiluminescence flow-sensing system for hydrogen peroxide with immobilized reagents

被引:47
作者
Qin, W
Zhang, ZJ [1 ]
Li, BX
Liu, SN
机构
[1] Shaanxi Normal Univ, Dept Chem, Xian 710062, Peoples R China
[2] Nanjing Univ, Dept Chem, Nanjing 210093, Peoples R China
基金
中国国家自然科学基金;
关键词
chemiluminescence; flow sensor; hydrogen peroxide; glucose;
D O I
10.1016/S0003-2670(98)00371-7
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel chemiluminescence (CL) sensing system for hydrogen peroxide combined with flow-injection analysis (FIA) is described in this paper. It is prepared by electrostatically immobilizing the analytical reagents, luminol and cobalt(II), on a strongly basic anion-exchange resin and a weakly acid cation-exchange resin, respectively. Hydrogen peroxide is sensed by the CL reaction with luminol and cobalt(II) bleeding from the ion-exchange column with immobilized reagents by hydrolysis. The calibration graph is linear in the range 4 x 10(-8) to 1 x 10(-5) mol l(-1), and the detection limit is 1.2 x 10(-8) mol l(-1) hydrogen peroxide. A complete analysis could be performed in 1 min with a relative standard deviation of <5%. The system could be used for 50 h and has been applied successfully to the determination of hydrogen peroxide in rainwater and glucose in serum by measuring the formation of hydrogen peroxide from a packed bed reactor with immobilized glucose oxidase. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:357 / 363
页数:7
相关论文
共 25 条
[1]   ONLINE CHEMILUMINESCENCE ASSAY USING FIA AND FIBER OPTICS FOR URINARY AND BLOOD-GLUCOSE [J].
CATTANEO, MV ;
LUONG, JHT .
ENZYME AND MICROBIAL TECHNOLOGY, 1993, 15 (05) :424-428
[2]  
CHARPENTIER L, 1995, ANALUSIS, V23, P265
[3]  
CI Y, 1989, CHINESE J ANAL CHEM, V17, P404
[4]  
FERNANDEZROMERO JM, 1993, ANAL CHEM, V65, P3046
[5]   RETENTION BEHAVIOR OF SOME AROMATIC-COMPOUNDS ON CHEMICALLY BONDED CYCLODEXTRIN SILICA STATIONARY PHASE IN LIQUID-CHROMATOGRAPHY [J].
FUJIMURA, K ;
UEDA, T ;
ANDO, T .
ANALYTICAL CHEMISTRY, 1983, 55 (03) :446-450
[6]   SPECTROPHOTOMETRIC DETERMINATION OF HYDROGEN-PEROXIDE USING O-HYDROXYHYDROQUINONEPHTHALEIN, TITANIUM(IV) AND ETHYLENEDIAMINETETRAACETIC ACID [J].
FUJITA, Y ;
MORI, I ;
TOYODA, M ;
MATSUO, T .
ANALYTICAL SCIENCES, 1994, 10 (05) :827-830
[7]   CHEMILUMINESCENCE ANALYSIS IN FLOWING STREAMS WITH LUMINOL IMMOBILIZED ON SILICA AND CONTROLLED-PORE GLASS [J].
HOOL, K ;
NIEMAN, TA .
ANALYTICAL CHEMISTRY, 1987, 59 (06) :869-872
[8]   FLUORIMETRIC DETERMINATION OF TRACE HYDROGEN-PEROXIDE IN WATER WITH A FLOW-INJECTION SYSTEM [J].
HWANG, H ;
DASGUPTA, PK .
ANALYTICA CHIMICA ACTA, 1985, 170 (APR) :347-352
[9]  
ISHII M, 1992, BUNSEKI KAGAKU, V41, P125
[10]   FLOW-INJECTION AND PHOTOMETRIC-DETERMINATION OF HYDROGEN-PEROXIDE IN RAINWATER WITH N-ETHYL-N-(SULFOPROPYL)ANILINE SODIUM-SALT [J].
MADSEN, BC ;
KROMIS, MS .
ANALYTICAL CHEMISTRY, 1984, 56 (14) :2849-2850