Joint Reconstruction of Image and Motion in Gated Positron Emission Tomography

被引:74
作者
Blume, Moritz [1 ,2 ]
Martinez-Moeller, Axel [3 ]
Keil, Andreas
Navab, Nassir
Rafecas, Magdalena [1 ]
机构
[1] Univ Valencia, CSIC, IFIC, E-46071 Valencia, Spain
[2] Tech Univ Munich, Dept Comp Sci, D-85748 Munich, Germany
[3] Tech Univ Munich, Klinikum Rechts Isar, Nukl Med Klin, D-81675 Munich, Germany
关键词
Gating; motion compensation; positron emission tomography (PET); reconstruction; RESPIRATORY MOTION; CORONARY-ARTERIES; PET;
D O I
10.1109/TMI.2010.2053212
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
We present a novel intrinsic method for joint reconstruction of both image and motion in positron emission tomography (PET). Intrinsic motion compensation methods exclusively work on the measured data, without any external motion measurements. Most of these methods separate image from motion estimation: They use deformable image registration/optical flow techniques in order to estimate the motion from individually reconstructed gates. Then, the image is estimated based on this motion information. With these methods, a main problem lies in the motion estimation step, which is based on the noisy gated frames. The more noise is present, the more inaccurate the image registration becomes. As we show both visually and quantitatively, joint reconstruction using a simple deformation field motion model can compete with state-of-the-art image registration methods which use robust multilevel B-spline motion models.
引用
收藏
页码:1892 / 1906
页数:15
相关论文
共 37 条
[1]  
[Anonymous], ISCAS 2001
[2]  
BLUME M, 2009, P SOC PHOTO-OPT INS, V7258, P72
[3]  
Blume M, 2008, IEEE NUCL SCI S OCT, P5485
[4]  
Boucher L, 2004, J NUCL MED, V45, P214
[5]   Three-dimensional motion estimation with image reconstruction for gated cardiac ECT [J].
Cao, ZX ;
Gilland, DR ;
Mari, BA ;
Jaszczak, RJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (03) :384-388
[6]   A simple fixed-point approach to invert a deformation field [J].
Chen, Mingli ;
Lu, Weiguo ;
Chen, Quan ;
Ruchala, Kenneth J. ;
Olivera, Gustavo H. .
MEDICAL PHYSICS, 2008, 35 (01) :81-88
[7]   SPATIAL RESOLUTION AND NOISE PROPERTIES OF REGULARIZED MOTION-COMPENSATED IMAGE RECONSTRUCTION [J].
Chun, Se Young ;
Fessler, Jeffrey A. .
2009 IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: FROM NANO TO MACRO, VOLS 1 AND 2, 2009, :863-866
[8]   Relationship between motion of coronary arteries and diaphragm during free breathing: Lessons from real-time MR imaging [J].
Danias, PG ;
Stuber, M ;
Botnar, RM ;
Kissinger, KV ;
Edelman, RR ;
Manning, WJ .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1999, 172 (04) :1061-1065
[9]   Lung motion correction on respiratory gated 3-D PET/CT images [J].
Dawood, M ;
Lang, N ;
Jiang, XY ;
Schäfers, KP .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2006, 25 (04) :476-485
[10]   Motion estimation for cardiac emission tomography by optical flow methods [J].
Gilland, D. R. ;
Mair, B. A. ;
Parker, J. G. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (11) :2991-3006