Estimation of respiratory volumes from the photoplethysmographic signal.: Part 2:: a model study

被引:45
作者
Johansson, A [1 ]
Öberg, PÅ [1 ]
机构
[1] Linkoping Univ, Dept Biomed Engn, S-58185 Linkoping, Sweden
关键词
photoplethysmography; respiratory volume; physiological modelling; noninvasive techniques;
D O I
10.1007/BF02513265
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
A Windkessel model has been constructed with the aim of investigating the respiratory-volume dependence of the photoplethysmographic (PPG) signal. Experimental studies show a correlation between respiratory volume and the peak-to-peak value of the respiratory-induced intensity variations (RIIV) in the PPG signal. The model compartments are organised in two closed chambers, representing the thorax and the abdomen, and in a peripheral part not directly influenced by respiration. Cardiac pulse and respiration are created by continuous adjustment of the pressures in the affected compartments. Together with the criteria for heart and venous valves, the model is based on a set of 17 differential equations. These equations are salved for varying thoracic and abdominal pressures corresponding to different respiratory volumes. Furthermore, a sensitivity analysis is performed to evaluate the properties of the model. The PPG signals are created as a combination of peripheral blood flow and pressure. From these signals, the respiratory synchronous parts are extracted and analysed. To study some important limitations of the model, respiratory type and rate are varied. From the simulations, it is possible to verify our earlier experimental results concerning the relationship between respiratory volume and the peak-to-peak value of the RIIV signal. An expected decrease in the amplitude of the respiratory signal with increased respiratory rate is also found, which is due to the lowpass characteristics of the vessel system. Variations in the relationship between thoracic and abdominal respiration also affect the RIIV signal. The simulations explain and verify what has been found previously in experimental studies.
引用
收藏
页码:48 / 53
页数:6
相关论文
共 25 条
[11]   PHOTO-ELECTRIC PLETHYSMOGRAPHY AS A MONITORING DEVICE IN ANESTHESIA - APPLICATION AND INTERPRETATION [J].
DORLAS, JC ;
NIJBOER, JA .
BRITISH JOURNAL OF ANAESTHESIA, 1985, 57 (05) :524-530
[12]  
ELDRUPJO.S, 1966, SURGERY, V59, P505
[13]   MATHEMATICAL ANALYSIS AND DIGITAL SIMULATION OF RESPIRATORY CONTROL SYSTEM [J].
GRODINS, FS ;
BUELL, J ;
BART, AJ .
JOURNAL OF APPLIED PHYSIOLOGY, 1967, 22 (02) :260-&
[14]  
Hertzman AB, 1937, P SOC EXP BIOL MED, V37, P529
[15]   Estimation of respiratory volumes from the photoplethysmographic signal.: Part I:: experimental results [J].
Johansson, A ;
Öberg, PÅ .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1999, 37 (01) :42-47
[16]   MONITORING OF RESPIRATORY AND HEART-RATES USING A FIBEROPTIC SENSOR [J].
LINDBERG, LG ;
UGNELL, H ;
OBERG, PA .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1992, 30 (05) :533-537
[17]   PHYSICAL PROPERTIES OF HUMAN LUNGS MEASURED DURING SPONTANEOUS RESPIRATION [J].
MEAD, J ;
WHITTENBERGER, JL .
JOURNAL OF APPLIED PHYSIOLOGY, 1953, 5 (12) :779-796
[18]   RESPIRATORY REGULATION OF SPLANCHNIC AND SYSTEMIC VENOUS RETURN [J].
MORENO, AH ;
BURCHELL, AR ;
VANDERWO.R ;
BURKE, JH .
AMERICAN JOURNAL OF PHYSIOLOGY, 1967, 213 (02) :455-+
[19]  
Nitzan M, 1996, J Biomed Opt, V1, P223, DOI 10.1117/12.231366
[20]  
Noordergraaf A., 1978, CIRCULATORY SYSTEM D