Flow. in a mechanical bileaflet heart valve at laminar and near-peak systole flow rates: CFD simulations and experiments

被引:86
作者
Ge, L
Leo, HL
Sotiropoulos, F
Yoganathan, AP
机构
[1] Georgia Inst Technol, Walter H Coulter Sch Biomed Engn, Atlanta, GA 30332 USA
[2] Georgia Inst Technol, Sch Civil & Environm Engn, Atlanta, GA 30332 USA
来源
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME | 2005年 / 127卷 / 05期
关键词
D O I
10.1115/1.1993665
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Time accurate, fully 3D numerical simulations and particle image velocity laboratory experiments are carried out for flow through a fully open bileaflet mechanical heart valve under steady (nonpulsatile) inflow conditions. Flows at two different Reynolds numbers, one in the laminar regime and the other turbulent (near-peak systole flow rate), are investigated. A direct numerical simulation is carried out for the laminar flow ease while the turbulent flow is investigated with two different unsteady statistical turbulence modeling approaches, unsteady Reynolds-averaged Navier-Stokes (URANS) and detached-eddy simulation (DES) approach. For both the laminar and turbulent cases the computed mean velocity profiles are in good overall agreement. with the measurements. For the turbulent simulations, however, the comparisons with the measurements demonstrate clearly the superiority of the DES approach and underscore its potential as a powerful modeling tool of cardiovascular flows at physiological conditions. The study reveals numerous previously unknown features of the flow.
引用
收藏
页码:782 / 797
页数:16
相关论文
共 32 条
  • [1] Free emboli formation in the wake of bi-leaflet mechanical heart valves and the effects of implantation techniques
    Bluestein, D
    Li, YM
    Krukenkamp, IB
    [J]. JOURNAL OF BIOMECHANICS, 2002, 35 (12) : 1533 - 1540
  • [2] Experimental investigation of the steady flow downstream of the St. Jude bileaflet heart valve: A comparison between laser Doppler velocimetry and particle image velocimetry techniques
    Browne, P
    Ramuzat, A
    Saxena, R
    Yoganathan, AP
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 2000, 28 (01) : 39 - 47
  • [3] LASER ANEMOMETRY MEASUREMENTS OF PULSATILE FLOW PAST AORTIC-VALVE PROSTHESES
    CHANDRAN, KB
    CABELL, GN
    KHALIGHI, B
    CHEN, CJ
    [J]. JOURNAL OF BIOMECHANICS, 1983, 16 (10) : 865 - 873
  • [4] AN INVITRO COMPARATIVE-STUDY OF ST-JUDE MEDICAL AND EDWARDS-DUROMEDICS BILEAFLET VALVES USING LASER ANEMOMETRY
    FATEMI, R
    CHANDRAN, KB
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1989, 111 (04): : 298 - 302
  • [5] ON THE HEMOLYTIC AND THROMBOGENIC POTENTIAL OF OCCLUDER PROSTHETIC HEART-VALVES FROM INVITRO MEASUREMENTS
    FIGLIOLA, RS
    MUELLER, TJ
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1981, 103 (02): : 83 - 90
  • [6] Verification and validation studies of the time-averaged velocity field in the very near-wake of a finite elliptical cylinder
    Flynn, MR
    Eisner, AD
    [J]. FLUID DYNAMICS RESEARCH, 2004, 34 (04) : 273 - 288
  • [7] Fontaine AA, 1996, J HEART VALVE DIS, V5, P186
  • [8] Numerical simulation of flow in mechanical heart valves: Grid resolution and the assumption of flow symmetry
    Ge, L
    Jones, SC
    Sotiropoulos, F
    Healy, TM
    Yoganathan, AP
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2003, 125 (05): : 709 - 718
  • [9] GE L, 2004, THESIS GEORGIA I TEC
  • [10] Flow control with noncircular jets
    Gutmark, EJ
    Grinstein, FF
    [J]. ANNUAL REVIEW OF FLUID MECHANICS, 1999, 31 : 239 - +