Comparison of factors that influence the measurement of corneal hysteresis in vivo and in vitro

被引:20
作者
Alhamad, Tariq A. [1 ]
Meek, Keith M. [1 ]
机构
[1] Cardiff Univ, Sch Optometry & Vis Sci, Struct Biophys Grp, Cardiff CF24 4LU, S Glam, Wales
基金
英国医学研究理事会;
关键词
corneal hysteresis; in vitro; ocular response analyser; OCULAR RESPONSE ANALYZER; OPEN-ANGLE GLAUCOMA; BIOMECHANICAL PROPERTIES; INTRAOCULAR-PRESSURE; NORMAL EYES; EXTENSIBILITY; CHILDREN; INTACT; AGE;
D O I
10.1111/j.1755-3768.2011.02150.x
中图分类号
R77 [眼科学];
学科分类号
100212 [眼科学];
摘要
Purpose: The purpose of this study was to compare measurements of corneal hysteresis (CH) obtained in vivo, with similar measurements from excised human eyes and from excised human corneas mounted in an artificial anterior chamber. Methods: Corneal hysteresis was measured using an ocular response analyser (Reichert Ophthalmic Instruments) from three groups: 53 healthy normal corneas of fifty-three patients, six excised eyes and 17 excised corneas. Results: In vivo, it was found that CH was independent of gender, age and mean spherical equivalent, but has a significant inverse relationship with intraocular pressure (IOPcc) (r = 0.53; p < 0.0001). However, there was no correlation between CH and IOPG (r = 0.10; p = 461). The same inverse relationship with IOPcc was recorded in intact, excised eyes (r = 0.74; p < 0.0001), with no significant differences between the behaviour each individual eye. Excised corneas also showed an inverse relationship between CH and trans-corneal pressure (r = 0.72; p < 0.0001), but the measured values of CH were lower than those recorded in vivo and from intact globes. In both excised eyes and excised corneas, we found a significant correlation between CH and central corneal thickness [r = 0.86; p < 0.0001 and r = 0.611; p < 0.0005 (respectively)]. Conclusion: The in vitro results indicate that every normal human eye at physiological hydration shows an identical dependence of CH on IOPcc, the same dependence as is observed in vivo. This therefore would appear to be an intrinsic response of the tissue to a change in IOP. However, it is possible that the lower values of CH recorded from excised corneas reflect the influence of the artificial chamber replacing the eye globe, so in vivo values of CH may be influenced to some extent by the presence of the other components of the eye.
引用
收藏
页码:E443 / E450
页数:8
相关论文
共 35 条
[1]
Effect of central corneal thickness and corneal hysteresis on tonometry as measured by dynamic contour tonometry, Ocular Response Analyzer, and Goldmann tonometry in glaucomatous eyes [J].
Annette, Hager ;
Kristina, Loge ;
Bernd, Schroeder ;
Mark-Oliver, Fuellhas ;
Wolfgang, Wiegand .
JOURNAL OF GLAUCOMA, 2008, 17 (05) :361-365
[2]
Modelling the elastic properties of the anterior eye and their contribution to maintenance of image quality:: the role of the limbus [J].
Asejczyk-Widlicka, M. ;
Srodka, D. W. ;
Kasprzak, H. ;
Pierscionek, B. K. .
EYE, 2007, 21 (08) :1087-1094
[3]
Full-field deformation of bovine cornea under constrained inflation conditions [J].
Boyce, Brad L. ;
Grazier, J. Mark ;
Jones, Reese E. ;
Nguyen, Thao D. .
BIOMATERIALS, 2008, 29 (28) :3896-3904
[4]
Correlation between corneal hysteresis, corneal resistance factor, and ocular pulse amplitude in healthy subjects [J].
Ehongo, A. ;
de Maertelaer, V. ;
Cullus, P. ;
Pourjavan, S. .
JOURNAL FRANCAIS D OPHTALMOLOGIE, 2008, 31 (10) :999-1005
[5]
Assessment of corneal biomechanical properties and their variation with age [J].
Elsheikh, Ahmed ;
Wang, Defu ;
Brown, Michael ;
Rama, Paolo ;
Campanelli, Marino ;
Pye, David .
CURRENT EYE RESEARCH, 2007, 32 (01) :11-19
[6]
Assessment of the Ocular Response Analyzer as a Tool for Intraocular Pressure Measurement [J].
Elsheikh, Ahmed ;
Alhasso, Daad ;
Kotecha, Aachal ;
Garway-Heath, David .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2009, 131 (08)
[7]
Corneal Biomechanical Metrics in Eyes With Refraction of-19-00 to+9.00 D in Healthy Brazilian Patients [J].
Fontes, Bruno M. ;
Ambrosio, Renato ;
Alonso, Ruiz S. ;
Jardim, Daniela ;
Velarde, Guillermo C. ;
Nose, Walton .
JOURNAL OF REFRACTIVE SURGERY, 2008, 24 (09) :941-945
[8]
A viscoelastic biomechanical model of the cornea describing the effect of viscosity and elasticity on hysteresis [J].
Glass, Dianne H. ;
Roberts, Cynthia J. ;
Litsky, Alan S. ;
Weber, Paul A. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2008, 49 (09) :3919-3926
[9]
HJORTDAL JO, 1995, ACTA OPHTHALMOL SCAN, V73, P364
[10]
HJORTDAL JO, 1995, ACTA OPHTHALMOL SCAN, V73, P12