Right ventricular volume measurement by conductance catheter

被引:34
作者
Danton, MHD
Greil, GF
Byrne, JG
Hsin, M
Cohn, L
Maier, SE
机构
[1] Brigham & Womens Hosp, Dept Cardiac Surg, Boston, MA 02115 USA
[2] Brigham & Womens Hosp, Dept Radiol, Boston, MA 02115 USA
[3] Harvard Univ, Sch Med, Boston Childrens Hosp, Dept Cardiol, Boston, MA 02115 USA
来源
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY | 2003年 / 285卷 / 04期
关键词
magnetic resonance imaging; conductance catheter;
D O I
10.1152/ajpheart.00048.2003
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Continuous ventricular volume measurement by the conductance method assumes a homogeneous electrical field dispersed throughout and contained within the ventricle. Because of dense trabeculation and complex geometry, right ventricular ( RV) volume description by this method may be seriously compromised. This study sought to determine the accuracy and limitations of RV volume measurement by conductance, with magnetic resonance (MR) imaging (MRI) used as a reference, in the porcine RV. Anesthetized pigs ( n = 5, 45 - 55 kg) were placed in a 1.5-T magnet, and ECG-gated transverse MR images ( 5-mm slices) were acquired during the complete cardiac cycle. RV cavity volumes were subsequently determined by Simpson's technique. Animals were then instrumented with an RV conductance catheter and an ultrasonic pulmonary artery flow probe. Conductance catheter signals were recorded using single- and dual-field ( SF and DF) excitation, and the saline-dilution technique was used to correct volumes for parallel conductance. The gain factor (alpha) was calculated as the ratio of conductance- to MRI-derived stroke volume (alpha(SV)). Variation of alpha during the cardiac cycle was computed by comparing RV conductance volumes with 1) MRI volumes at isochronal time points within the cardiac cycle [alpha( t)] and 2) the pulmonary flow integral during ejection. After calibration, the conductance- MRI volume relation was modeled linearly with good correlation [ r = 0.96 ( SF) and r = 0.94 ( DF)], close to the line of identity. Individual conductance- MRI plots displayed a slight curvilinear relation that was concave toward the MRI axis. Consistent with this finding, alpha( t) varied significantly during the cardiac cycle (0.49 and 0.39 by SF for end systole and end diastole, respectively, P = 0.011). DF excitation resulted in improved volume measurement [alpha(SV) = 0.41 ( SF) and 0.96 ( DF)], with less variation in alpha( t) (1.0 and 0.92 by DF for end systole and end diastole, respectively, P = 0.66). These results indicate that, with calibration, the conductance method can measure absolute RV volume under steady-state conditions. However, the curvilinearity and alpha( t) variation would indicate the potential for nonlinearity when RV volumes are varied over a wider range.
引用
收藏
页码:H1774 / H1785
页数:12
相关论文
共 34 条
[1]  
AMIRHAMZEH MMR, 1994, CIRCULATION, V90, P347
[2]   SIMULTANEOUS CONDUCTANCE CATHETER AND DIMENSION ASSESSMENT OF LEFT-VENTRICLE VOLUME IN THE INTACT ANIMAL [J].
APPLEGATE, RJ ;
CHENG, CP ;
LITTLE, WC .
CIRCULATION, 1990, 81 (02) :638-648
[3]   CONTINUOUS MEASUREMENT OF LEFT-VENTRICULAR VOLUME IN ANIMALS AND HUMANS BY CONDUCTANCE CATHETER [J].
BAAN, J ;
VANDERVELDE, ET ;
DEBRUIN, HG ;
SMEENK, GJ ;
KOOPS, J ;
VANDIJK, AD ;
TEMMERMAN, D ;
SENDEN, J ;
BUIS, B .
CIRCULATION, 1984, 70 (05) :812-823
[4]  
Baan J., 1989, Automedica, V11, P357
[5]   CONTINUOUS STROKE VOLUME AND CARDIAC-OUTPUT FROM INTRAVENTRICULAR DIMENSIONS OBTAINED WITH IMPEDANCE CATHETER [J].
BAAN, J ;
JONG, TTA ;
KERKHOF, PLM ;
MOENE, RJ ;
VANDIJK, AD ;
VANDERVELDE, ET ;
KOOPS, J .
CARDIOVASCULAR RESEARCH, 1981, 15 (06) :328-334
[6]   LEFT-VENTRICULAR VOLUME MEASUREMENT BY CONDUCTANCE CATHETER IN INTACT DOGS - PARALLEL CONDUCTANCE VOLUME DEPENDS ON LEFT-VENTRICULAR SIZE [J].
BOLTWOOD, CM ;
APPLEYARD, RF ;
GLANTZ, SA .
CIRCULATION, 1989, 80 (05) :1360-1377
[7]   ACCURACY OF VOLUME MEASUREMENT BY CONDUCTANCE CATHETER IN ISOLATED, EJECTING CANINE HEARTS [J].
BURKHOFF, D ;
VANDERVELDE, E ;
KASS, D ;
BAAN, J ;
MAUGHAN, WL ;
SAGAWA, K .
CIRCULATION, 1985, 72 (02) :440-447
[8]   Right ventricular volume measurement using the conductance catheter method: Validation in excised porcine hearts [J].
Danton, MHD ;
Byrne, JG ;
Hsin, M ;
Laurence, R ;
Cohn, LH ;
Aklog, L .
ASAIO JOURNAL, 2002, 48 (05) :514-519
[9]  
Gopakumaran Bala, 1996, Biomedical Instrumentation and Technology, V30, P427
[10]   Measurement of left ventricular volume by dual-field conductance catheter in humans - Comparison with single-field conductance catheter [J].
Hayashi, Y ;
Takeuchi, M ;
Takaoka, H ;
Yokoyama, M .
JAPANESE CIRCULATION JOURNAL-ENGLISH EDITION, 1996, 60 (02) :85-95