A novel biosensor for specific determination of hydrogen peroxide:: catalase enzyme electrode based on dissolved oxygen probe

被引:36
作者
Akgöl, S [1 ]
Dinçkaya, E [1 ]
机构
[1] Ege Univ, Fac Sci, Dept Biochem, TR-35100 Bornova Izmir, Turkey
关键词
catalase; hydrogen peroxide; biosensor; catalase electrode; catalase immobilization;
D O I
10.1016/S0039-9140(98)00255-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A biosensor for the specific determination of hydrogen peroxide was developed using catalase (EC 1.11.1.6) in combination with a dissolved oxygen probe. Catalase was immobilized with gelatin by means of glutaraldehyde and fixed on a pretreated teflon membrane served as enzyme electrode. The electrode response was maximum when 50 mM phosphate buffer was used at pH 7.0 and at 35 degrees C. The biosensor response depends linearly on hydrogen peroxide concentration between 1.0 x 10(-5) and 3.0 x 10(-3) M with a. response time of 30 s. The sensor is stable for >3 months so in this period >400 assays can be performed. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:363 / 367
页数:5
相关论文
共 15 条
[1]   INDIRECT SPECTROPHOTOMETRIC DETERMINATION OF HYDROGEN-PEROXIDE [J].
AFSAR, H ;
DEMIRATA, B .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1993, 347 (10-11) :460-461
[2]  
CIUCU A, 1985, ANAL LETT PT B, V18, P299, DOI 10.1080/00032718508066133
[3]   CHEMICALLY IMMOBILIZED ENZYME ELECTRODES FOR HYDROGEN-PEROXIDE DETERMINATION [J].
COSGROVE, M ;
MOODY, GJ ;
THOMAS, JDR .
ANALYST, 1988, 113 (12) :1811-1815
[4]  
DAJUN C, 1993, SENSOR ACTUAT B-CHEM, V13, P554
[5]  
DEMMANO G, 1996, EUR J BIOCHEM, V238, P735
[6]  
DINCKAYA E, 1993, INDIAN J BIOCHEM BIO, V30, P282
[7]  
DINCKAYA E, 1994, TURK J CHEM, V18, P1
[8]   A SIMPLIFIED ENZYME-BASED FIBER OPTIC SENSOR FOR HYDROGEN-PEROXIDE AND OXIDASE SUBSTRATES [J].
LI, JZ ;
ZHANG, ZJ ;
LI, L .
TALANTA, 1994, 41 (11) :1999-2002
[9]   INFLUENCE OF GRAPHITE POWDER, ADDITIVES AND ENZYME IMMOBILIZATION PROCEDURES ON A MEDIATORLESS HRP-MODIFIED CARBON-PASTE ELECTRODE FOR AMPEROMETRIC FLOW-INJECTION DETECTION OF H2O2 [J].
POPESCU, IC ;
ZETTERBERG, G ;
GORTON, L .
BIOSENSORS & BIOELECTRONICS, 1995, 10 (05) :443-461
[10]   PHOTODIODE-BASED CHEMILUMINOMETRIC BIOSENSORS FOR HYDROGEN-PEROXIDE AND L-LYSINE [J].
PREUSCHOFF, F ;
SPOHN, U ;
JANASEK, D ;
WEBER, E .
BIOSENSORS & BIOELECTRONICS, 1994, 9 (08) :543-549