The TALON removable head frame system for stereotactic radiosurgery/radiotherapy: Measurement of the repositioning accuracy

被引:52
作者
Salter, BJ
Fuss, M
Vollmer, DG
Sadeghi, A
Bogaev, CA
Cheek, DA
Herman, TS
Hevezi, JM
机构
[1] Univ Texas, Hlth Sci Ctr, Dept Radiat Oncol, San Antonio, TX 78229 USA
[2] Univ Texas, Hlth Sci Ctr, Dept Neurosurg, Canc Therapy & Res Ctr, San Antonio, TX 78229 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2001年 / 51卷 / 02期
关键词
stereotactic radiotherapy; IMRT; immobilization; fixation; repositioning accuracy;
D O I
10.1016/S0360-3016(01)01670-4
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To present the TALON removable head frame system as an immobilization device for single-fraction intensity-modulated stereotactic radiosurgery (IMRS) and fractionated stereotactic intensity-modulated radiotherapy (FS-IMRT); and to evaluate the repositioning accuracy by measurement of anatomic landmark coordinates in repeated computed tomography (CT) examinations. Methods and Materials: Nine patients treated by fractionated stereotactic intensity-modulated radiotherapy underwent repeated CTs during their treatment courses. We evaluated anatomic landmark coordinates in a total of 26 repeat CT data sets and respective x, y, and z shifts relative to their positions in the nine treatment-planning reference CTs. An iterative optimization algorithm was employed using a root mean square scoring function to determine the best-fit orientation of subsequent sets of anatomic landmark measurements relative to the original image set. This allowed for the calculation or the x, y, and z components of translation of the target isocenter for each repeat CT. In addition to absolute target isocenter translation, the magnitude (sum vector) of isocenter motion and the patient/target rotation about the three principal axes were calculated. Results: Anatomic landmark analysis over a treatment course of 6 weeks revealed a mean target isocenter translation of 0.95 +/- 0.55, 0.58 +/- 0.46, and 0.51 +/- 0.38 mm in x, y, and z directions, respectively. The mean magnitude of isocenter translation was 1.38 +/- 0.48 mm. The 95% confidence interval ([CI], mean translation plus two standard deviations) for repeated isocenter setup accuracy over the 6-week period was 2.34 mm. Average rotations about the x, y. and z axes were 0.41 +/- 0.36, 0.29 +/- 0.25, and 0.18 +/- 0.15 degrees, respectively. Analysis of the accuracy of the first repeated setup control, representative of single-fraction stereotactic radiosurgery situations, resulted in a mean target isocenter translation in the x, y, and z directions of 0.52 +/- 0.38, 0.56 +/- 0.30, and 0.46 +/- 0.25 mm, respectively. The mean magnitude of isocenter translation was 0.99 +/- 0.28 mm. The 95% confidence interval for these radiosurgery situations was 1.55 mm. Average rotations at first repeated setup control about the x, y, and z axes were 0.24 +/- 0.19, 0.19 +/- 0.17, and 0.19 +/- 0.12 degrees, respectively. Conclusion: The TALON relocatable head frame was seen to be well suited for immobilization and repositioning of single fraction stereotactic radiosurgery treatments. Because of its unique removable design, the system was also seen to provide excellent repeat immobilization and alignment for fractionated stereotactic applications. The exceptional accuracy for the single-fraction stereotactic radiosurgical application of the system was seen to deteriorate only slightly over a 6-week fractionated stereotactic treatment course. (C) 2001 Elsevier Science Inc.
引用
收藏
页码:555 / 562
页数:8
相关论文
共 14 条
[1]   Computer-assisted neurosurgery by using a noninvasive vacuum-affixed dental cast that acts as a reference base: another step toward a unified approach in the treatment of brain tumors [J].
Bale, RJ ;
Burtscher, J ;
Eisner, W ;
Obwegeser, AA ;
Rieger, M ;
Sweeney, RA ;
Dessl, A ;
Giacomuzzi, SM ;
Twerdy, K ;
Jaschke, W .
JOURNAL OF NEUROSURGERY, 2000, 93 (02) :208-213
[2]   The university of Florida frameless high-precision stereotactic radiotherapy system [J].
Bova, FJ ;
Buatti, JM ;
Friedman, WA ;
Mendenhall, WM ;
Yang, CC ;
Liu, C .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (04) :875-882
[3]   HALO RING SUPPORTING THE BROWN-ROBERTS-WELLS STEREOTAXIC FRAME FOR FRACTIONATED RADIOTHERAPY [J].
CARINI, S ;
SCIELZO, G ;
RUGGIERI, FG ;
BISTOLFI, F ;
RAVEGNANI, M ;
ANDREUSSI, L .
ACTA NEUROCHIRURGICA, 1994, 129 (1-2) :92-96
[4]   A HALO-RING TECHNIQUE FOR FRACTIONATED STEREOTAXIC RADIOTHERAPY [J].
CLARK, BG ;
PODGORSAK, EB ;
SOUHAMI, L ;
OLIVIER, A ;
SIXEL, KE ;
CARON, JL .
BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (786) :522-527
[5]   FRACTIONATED RADIOTHERAPY OF SMALL INOPERABLE LESIONS OF THE BRAIN USING A NONINVASIVE STEREOTAXIC FRAME [J].
DELANNES, M ;
DALY, NJ ;
BONNET, J ;
SABATIER, J ;
TREMOULET, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (03) :749-755
[6]   Repositioning accuracy of a noninvasive head fixation system for stereotactic radiotherapy [J].
Hamilton, RJ ;
Kuchnir, FT ;
Pelizzari, CA ;
Sweeney, PJ ;
Rubin, SJ .
MEDICAL PHYSICS, 1996, 23 (11) :1909-1917
[7]   Three-dimensional accuracy and interfractional reproducibility of patient fixation and positioning using a stereotactic head mask system [J].
Karger, CP ;
Jäkel, O ;
Debus, J ;
Kuhn, S ;
Hartmann, GH .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 49 (05) :1493-1504
[8]   ADAPTATION AND VERIFICATION OF THE RELOCATABLE GILL-THOMAS-COSMAN FRAME IN STEREOTAXIC RADIOTHERAPY [J].
KOOY, HM ;
DUNBAR, SF ;
TARBELL, NJ ;
MANNARINO, E ;
FERARRO, N ;
SHUSTERMAN, S ;
BELLERIVE, M ;
FINN, L ;
MCDONOUGH, CV ;
LOEFFLER, JS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 30 (03) :685-691
[9]   COMPUTER-SYSTEMS AND MECHANICAL TOOLS FOR STEREOTAXICALLY GUIDED CONFORMATION THERAPY WITH LINEAR ACCELERATORS [J].
SCHLEGEL, W ;
PASTYR, O ;
BORTFELD, T ;
BECKER, G ;
SCHAD, L ;
GADEMANN, G ;
LORENZ, WJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1992, 24 (04) :781-787
[10]   SMALL-FIELD STEREOTAXIC EXTERNAL-BEAM RADIATION-THERAPY OF INTRACRANIAL LESIONS - FRACTIONATED TREATMENT WITH A FIXED-HALO IMMOBILIZATION DEVICE [J].
SCHWADE, JG ;
HOUDEK, PV ;
LANDY, HJ ;
BUJNOSKI, JL ;
LEWIN, AA ;
ABITOL, AA ;
SERAGO, CF ;
PISCIOTTA, VJ .
RADIOLOGY, 1990, 176 (02) :563-565