Quality assurance of a system for improved target localization and patient set-up that combines real-time infrared tracking and stereoscopic X-ray imaging

被引:128
作者
Verellen, D [1 ]
Soete, G [1 ]
Linthout, N [1 ]
Van Acker, S [1 ]
De Roover, P [1 ]
Vinh-Hung, V [1 ]
Van de Steene, J [1 ]
Storme, G [1 ]
机构
[1] Free Univ Brussels, Acad Hosp, Dept Radiotherapy, Ctr Oncol, B-1090 Brussels, Belgium
关键词
automated patient positioning; intra-fractional correction of three-dimensional set-up errors; infrared markers; stereoscopic X-ray imaging;
D O I
10.1016/S0167-8140(02)00385-7
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: The aim of this study is to investigate the positional accuracy of a prototype X-ray imaging tool in combination with a real-time infrared tracking device allowing automated patient set-up in three dimensions. Material and methods: A prototype X-ray imaging tool has been integrated with a commercially released real-time infrared tracking device. The system, consisting of two X-ray tubes mounted to the ceiling and a centrally located amorphous silicon detector has been developed for automated patient positioning from outside the treatment room prior to treatment. Two major functions are supported: (a) automated fusion of the actual treatment images with digitally reconstructed radiographs (DRRs) representing the desired position; (b) matching of implanted radio opaque markers. Measurements of known translational (up to 30.0 mm) and rotational (up to 4.0degrees) set-up errors in three dimensions as well as hidden target tests have been performed on anthropomorphic phantoms. Results: The system's accuracy can be represented with the mean three-dimensional displacement vector, which yielded 0.6 mm (with an overall SD of 0.9 mm) for the fusion of DRRs and X-ray images. Average deviations between known translational errors and calculations varied from - 0.3 to 0.6 mm with a standard deviation in the range of 0.6-1.2 mm. The marker matching algorithm yielded a three-dimensional uncertainty of 0.3 rum (overall SD: 0.4 mm), with averages ranging from 0.0 to 0.3 mm and a standard deviation in the range between 0.3 and 0.4 turn. Conclusions: The stereoscopic X-ray imaging device integrated with the real-time infrared tracking device represents a positioning tool allowing for the geometrical accuracy that is required for conformal radiation therapy of abdominal and pelvic lesions, within an acceptable time-frame. (C) 2002 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:129 / 141
页数:13
相关论文
共 47 条
[31]  
SCHELL MC, 1995, 42 AAPM TASK GROUP
[32]  
SCHEWE JE, 1995, MED PHYS, V22, P939
[33]   Four-dimensional treatment planning and fluoroscopic real-time tumor tracking radiotherapy for moving tumor [J].
Shirato, H ;
Shimizu, S ;
Kitamura, K ;
Nishioka, T ;
Kagei, K ;
Hashimoto, S ;
Aoyama, H ;
Kunieda, T ;
Shinohara, N ;
Dosaka-Akita, H ;
Miyasaka, K .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 48 (02) :435-442
[34]   TECHNIQUE FOR VERIFYING TREATMENT FIELDS USING PORTAL IMAGES WITH DIAGNOSTIC QUALITY [J].
SHIU, AS ;
HOGSTROM, KR ;
JANJAN, NA ;
FIELDS, RS ;
PETERS, LJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1987, 13 (10) :1589-1594
[35]   Initial clinical experience with infrared-reflecting skin markers in the positioning of patients treated by conformal radiotherapy for prostate cancer [J].
Soete, G ;
Van de Steene, J ;
Verellen, D ;
Vinh-Hung, V ;
Van den Berge, D ;
Michielsen, D ;
Keuppens, F ;
De Roover, P ;
Storme, G .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 52 (03) :694-698
[36]   Clinical use of stereoscopic X-ray positioning of patients treated with conformal radiotherapy for prostate cancer [J].
Soete, G ;
Verellen, D ;
Michielsen, D ;
Vinh-Hung, V ;
Van de Steene, J ;
Van den Berge, D ;
De Roover, P ;
Keuppens, F ;
Storme, G .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 54 (03) :948-952
[37]   Electronic portal imaging with on-line correction of setup error in thoracic irradiation: Clinical evaluation [J].
Van de Steene, J ;
Van den Heuvel, F ;
Bel, A ;
Verellen, D ;
De Mey, J ;
Noppen, M ;
De Beukeleer, M ;
Storme, G .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 40 (04) :967-976
[38]   Microprocessor controlled limitation system for a stand-alone freely movable treatment couch [J].
Van de Vondel, I ;
Coppens, L ;
Verellen, D ;
Bel, A ;
Linthout, N ;
De Beukeleer, M ;
Van den Berge, D ;
Storme, G .
MEDICAL PHYSICS, 1998, 25 (06) :897-899
[39]   Remote control for a stand-alone freely movable treatment couch with limitation system [J].
Van de Vondel, I ;
Coppens, L ;
Verellen, D ;
Linthout, N ;
Van Acker, S ;
Storme, G .
MEDICAL PHYSICS, 2001, 28 (12) :2518-2521
[40]   CLINICAL IMPLEMENTATION OF AN OBJECTIVE COMPUTER-AIDED PROTOCOL FOR INTERVENTION IN INTRA-TREATMENT CORRECTION USING ELECTRONIC PORTAL IMAGING [J].
VANDENHEUVEL, F ;
DENEVE, W ;
VERELLEN, D ;
COGHE, M ;
COEN, V ;
STORME, G .
RADIOTHERAPY AND ONCOLOGY, 1995, 35 (03) :232-239