Analysis and reduction of 3D systematic and random setup errors during the simulation and treatment of lung cancer patients with CT-based external beam radiotherapy dose planning

被引:92
作者
de Boer, HCJ
De Koste, JRV
Senan, S
Visser, AG
Heijmen, BJM
机构
[1] Univ Rotterdam Hosp, Dr Daniel Den Hoed Canc Ctr, Div Clin Phys, Dept Radiat Oncol, NL-3008 AE Rotterdam, Netherlands
[2] Univ Nijmegen, Joint Ctr Radiat Oncol Arnhem Nijmegen, Nijmegen, Netherlands
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2001年 / 49卷 / 03期
关键词
setup accuracy; lung cancer; treatment simulation; correction protocol; EPID;
D O I
10.1016/S0360-3016(00)01413-9
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To determine the magnitude of the errors made in (a) the setup of patients with lung cancer on the simulator relative to their intended setup with respect to the planned treatment beams and (b) in the setup of these patients on the treatment unit, To investigate how the systematic component of the latter errors can be reduced with an off-line derision protocol for setup corrections. Methods and Materials: For 39 patients with CT planning, digitally-reconstructed radiographs (DRRs) were calculated for anterior-posterior and lateral beams. Retrospectively, the position of the visible anatomy relative to the planned isocenter was compared with the corresponding position on the digitized simulator radiographs using contour match software. The setup accuracy at the treatment unit relative to the simulator setup was measured for 40 patients for at least 5 fractions per patient in 2 orthogonal beams with the aid of an electronic portal imaging device (EPID), Setup corrections were applied, based on an off-line decision protocol, with parameters derived from knowledge of the random setup errors in the studied patient group. Results: The standard deviations (SD) of the simulator setup errors relative to the CT planning setup in the lateral, longitudinal, and anterior-posterior directions were 4.0, 2.8, and 2.5 mm, respectively. The SD of rotations around the anterior-posterior axis was 1.6 degrees and around the left-right axis 1.3 degrees. The setup error at the treatment unit had a small random component in all three directions (1 SD = 2 mm), The systematic components were larger, particularly in the longitudinal direction (1 SD = 3.6 mm), but were reduced with the decision protocol to 1 SD < 2 mm with, on average, 0.6 setup correction per patient. Conclusion: Setup errors at the simulator, which become systematic errors if the simulation defines the reference setup, were comparable to the systematic setup errors at the treatment unit in case no off-line protocol would have been applied. Hence, the omission of a separate simulation step can reduce systematic errors as efficiently as the application of an off-line correction protocol during treatment. The random errors were sufficiently small to make an off-line protocol feasible, (C) 2001 Elsevier Science Inc.
引用
收藏
页码:857 / 868
页数:12
相关论文
共 28 条
[1]  
[Anonymous], RAD THERAPY PHYS
[2]   Uncertainties in CT-based radiation therapy treatment planning associated with patient breathing [J].
Balter, JM ;
TenHaken, RK ;
Lawrence, TS ;
Lam, KL ;
Robertson, JM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 36 (01) :167-174
[3]   High-precision prostate cancer irradiation by clinical application of an offline patient setup verification procedure, using portal imaging [J].
Bel, A ;
Vos, PH ;
Rodrigus, PTR ;
Creutzberg, CL ;
Visser, AG ;
Stroom, JC ;
Lebesque, JV .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 35 (02) :321-332
[4]   A VERIFICATION PROCEDURE TO IMPROVE PATIENT SET-UP ACCURACY USING PORTAL IMAGES [J].
BEL, A ;
VANHERK, M ;
BARTELINK, H ;
LEBESQUE, JV .
RADIOTHERAPY AND ONCOLOGY, 1993, 29 (02) :253-260
[5]   TRANSFER ERRORS OF PLANNING CT TO SIMULATOR - A POSSIBLE SOURCE OF SETUP INACCURACIES [J].
BEL, A ;
BARTELINK, H ;
VIJLBRIEF, RE ;
LEBESQUE, JV .
RADIOTHERAPY AND ONCOLOGY, 1994, 31 (02) :176-180
[6]   Impact of cradle immobilization on setup reproducibility during external beam radiation therapy for lung cancer [J].
Bentel, GC ;
Marks, LB ;
Krishnamurthy, R .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (03) :527-531
[7]   MAXIMIZING SETUP ACCURACY USING PORTAL IMAGES AS APPLIED TO A CONFORMAL BOOST TECHNIQUE FOR PROSTATIC-CANCER [J].
BIJHOLD, J ;
LEBESQUE, JV ;
HART, AAM ;
VIJLBRIEF, RE .
RADIOTHERAPY AND ONCOLOGY, 1992, 24 (04) :261-271
[8]   Response, toxicity, failure patterns, and survival in five radiation therapy oncology group (RTOG) trials of sequential and/or concurrent chemotherapy and radiotherapy for locally advanced non-small-cell carcinoma of the lung [J].
Byhardt, RW ;
Scott, C ;
Sause, WT ;
Emami, B ;
Komaki, R ;
Fisher, B ;
Lee, JS ;
Lawton, C .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 42 (03) :469-478
[9]  
CHEVALIER T, 1992, J NATL CANCER I, V184, P158
[10]   Characterization of a high-elbow, fluoroscopic electronic portal imaging device for portal dosimetry [J].
de Boer, JCJ ;
Heijmen, BJM ;
Pasma, KL ;
Visser, AG .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (01) :197-216