Interventional cardiovascular procedures guided by real-time MR imaging: An interactive interface using multiple slices, adaptive projection modes and live 3D renderings

被引:46
作者
Guttman, Michael A. [1 ]
Ozturk, Cenghizhan [2 ]
Raval, Amish N. [2 ]
Raman, Venkatesh K. [2 ]
Dick, Alexander J. [2 ]
DeSilva, Ranil [2 ]
Karmarkar, Parag [2 ]
Lederman, Robert J. [2 ]
McVeigh, Elliot R. [1 ]
机构
[1] NHLBI, NIH, Div Intramural Res, Cardiac Energet Lab, Bethesda, MD 20892 USA
[2] NHLBI, NIH, Div Intramural Res, Cardiovasc Branch, Bethesda, MD 20892 USA
关键词
interventional MRI; real-time MRI; image guided interventions; catheterization; user interface;
D O I
10.1002/jmri.21199
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To develop and test a novel interactive real-time MRI environment that facilitates image-guided cardiovascular interventions. Materials and Methods: Color highlighting of device-mounted receiver coils, accelerated imaging of multiple slices, adaptive projection modes, live three-dimensional (3D) renderings and other interactive features were utilized to enhance navigation of devices and targeting of tissue. Results: Images are shown from several catheter-based interventional procedures performed in swine that benefit from this custom interventional MRI interface. These include endograft repair of aortic aneurysm, balloon septostomy of the cardiac interatrial septum, angioplasty and stenting, and endomyocardial cell injection, all using active catheters containing MRI receiver coils. Conclusion: Interactive features not available on standard clinical scanners enhance real-time MRI for guiding cardiovascular interventional procedures.
引用
收藏
页码:1429 / 1435
页数:7
相关论文
共 34 条
[31]   Magnetization preparation during the steady state: Fat-saturated 3D TrueFISP [J].
Scheffler, K ;
Heid, O ;
Hennig, J .
MAGNETIC RESONANCE IN MEDICINE, 2001, 45 (06) :1075-1080
[32]  
Serfaty JM, 2000, J MAGN RESON IMAGING, V12, P590, DOI 10.1002/1522-2586(200010)12:4<590::AID-JMRI11>3.0.CO
[33]  
2-3
[34]   MR coil design for simultaneous tip tracking and curvature delineation of a catheter [J].
Zuehlsdorff, S ;
Umathum, R ;
Volz, S ;
Hallscheidt, P ;
Fink, C ;
Semmler, W ;
Bock, M .
MAGNETIC RESONANCE IN MEDICINE, 2004, 52 (01) :214-218