Wet and dry chemistry kits for total creatine kinase activity using a microfabricated, planar, small-volume, amperometric cell

被引:11
作者
Nagy, G
Xu, CX
Buck, RP
Lindner, E
Neuman, MR
Sprinkle, RH
机构
[1] Univ N Carolina, Dept Chem, Chapel Hill, NC 27599 USA
[2] Duke Univ, Dept Biomed Engn, Durham, NC 27708 USA
[3] Case Western Reserve Univ, Metro Hlth Med Ctr, Dept OB GYN, Cleveland, OH 44109 USA
[4] Univ Maryland, Sch Publ Affairs, College Pk, MD 20742 USA
基金
美国国家科学基金会;
关键词
D O I
10.1016/S0003-2670(98)00600-X
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel microfabricated amperometric microcell is used for total creatine kinase (CK) activity determination. The microcell employs a photolithograpically prepared flat form electrode pair for detection. The enzyme-catalyzed reaction takes place in a thin Liquid film, supported by a replaceable porous, hydrophillic membrane, resting on the working electrode surface. Simple, two electrode, amperometry was used, to avoid overvoltage shocks caused by "open" circuits. An electrochemically prepared size-exclusion membrane was utilized on the working electrode to provide selectivity for the amperometric detection of hydrogen peroxide. Two different analytical reaction sequences following the CK catalyzed creatine phosphate hydrolysis were investigated. Both result in hydrogen peroxide production. The rate of hydrogen peroxide production is proportional to the total CK activity, and it is followed after the dispensation of sample into the microcell. The slope of the amperometric current-time curve is evaluated as the analytical signal. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:1 / 12
页数:12
相关论文
共 12 条
[1]  
BLUM W, 1949, PRINCIPLES ELECTROPL, P382
[2]  
BURTIS CA, 1994, TXB CLIN CHEM, P802
[3]   USE OF POLYMER-FILMS IN AMPEROMETRIC BIOSENSORS [J].
EMR, SA ;
YACYNYCH, AM .
ELECTROANALYSIS, 1995, 7 (10) :913-923
[4]  
FINDLAY JB, 1983, AM ASS CLIN CHEM 35, V29, P1212
[5]  
GERHARDT W, 1983, METHOD ENZYMAT AN, V3, P508
[6]   RECOMMENDED METHOD FOR THE DETERMINATION OF CREATINE-KINASE IN BLOOD MODIFIED BY THE INCLUSION OF EDTA - COMMITTEE ON ENZYMES OF THE SCANDINAVIAN SOCIETY FOR CLINICAL-CHEMISTRY AND CLINICAL PHYSIOLOGY (SCE) [J].
HORDER, M ;
MAGID, E ;
PITKANEN, E ;
HARKONEN, M ;
STROMME, JH ;
THEODORSEN, L ;
GERHARDT, W ;
WALDENSTROM, J .
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1979, 39 (01) :1-5
[7]   Miniaturized biosensors employing electropolymerized permselective films and their use for creatinine assays in human serum [J].
Madaras, MB ;
Buck, RP .
ANALYTICAL CHEMISTRY, 1996, 68 (21) :3832-3839
[8]   Microfabricated amperometric creatine and creatinine biosensors [J].
Madaras, MB ;
Popescu, IC ;
Ufer, S ;
Buck, RP .
ANALYTICA CHIMICA ACTA, 1996, 319 (03) :335-345
[9]   Amperometric monitoring of lactate accumulation in rabbit ischemic myocardium [J].
Marzouk, SAM ;
Cosofret, VV ;
Buck, RP ;
Yang, H ;
Cascio, WE ;
Hassan, SSM .
TALANTA, 1997, 44 (09) :1527-1541
[10]   Amperometric microcell for enzyme activity measurements [J].
Nagy, G ;
Xu, CX ;
Buck, RP ;
Lindner, E ;
Neuman, MR .
ANALYTICAL CHEMISTRY, 1998, 70 (10) :2156-2162