Chronic performance of steroid-eluting epicardial leads in a growing pediatric population: A 10-year comparison

被引:32
作者
Horenstein, MS
Hakimi, M
Walters, H
Karpawich, PP
机构
[1] Wayne State Univ, Childrens Hosp Michigan, Sch Med,Dept Pediat, Div Cardiol, Detroit, MI 48201 USA
[2] Wayne State Univ, Childrens Hosp Michigan, Sch Med,Dept Pediat, Div Cardiovasc Surg, Detroit, MI 48201 USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2003年 / 26卷 / 07期
关键词
steroid-eluting leads; epicardial electrode; pediatrics;
D O I
10.1046/j.1460-9592.2003.t01-1-00212.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Patient size and congenital heart defects complicate pacemaker therapy in children favoring an initial epicardial approach. Steroid-eluting (SE) epicardial (EPI) leads maintain stable, low pacing thresholds in the short-term When compared to the nonsteroid (NSE) epicardial (EPI) leads. The purpose of this study was to evaluate chronic, 10-year performance of SE leads in growing children compared with NSE EPI leads implanted during the same time interval. From 1990 to 2000, 35 patients (age 1 month to 18 year, median 3 years), 28 with and 7 without congenital heart disease (CHD) received 51 SE leads: 27 ventricular and 24 atrial. NSE leads were implanted in 27 patients (age 1-28 years, median 8 years), 24 with and 3 without CHD: 27 ventricular and 1 atrial. Pacing lead threshold, impedance, and energy were measured at implant and during a 10-year follow-up. Unpaired t-test showed that impedance remained stable for all leads with lower mean values for the SE (376 55 vs 443 109 Q) (P = NS). The mean energy requirement for SE leads at 10 years (1.2 +/- 0.9 muJ) was significantly lower than for NSE (4.4 +/- 5.5 muJ) (P < 0.05). At 2.5-V output, chronic thresholds for SE leads did not significantly differ from implant values for atrial (0.08 vs 0.09 ms) or ventricular (0.08 vs 0.08 ms) sites. There were no differences in SE lead performances among patients with or without CHD. Fracture or dislodgement occurred in two SE (4%) and four NSE (14%) leads. SE outperform NSE EPI leads and show stable, chronic low thresholds over time in all growing children.
引用
收藏
页码:1467 / 1471
页数:5
相关论文
共 17 条
[1]  
Beaufort-Krol GCM, 1999, J THORAC CARDIOV SUR, V117, P523
[2]  
Cohen MI, 2001, CIRCULATION, V103, P2585
[3]   Steroid-eluting epicardial pacing electrodes: Six year experience of pacing thresholds in a growing pediatric population [J].
Cutler, NG ;
Karpawich, PP ;
Cavitt, D ;
Hakimi, M ;
Walters, HL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (12) :2943-2948
[4]   PERMANENT EPICARDIAL AND TRANSVENOUS SINGLE-CHAMBER AND DUAL-CHAMBER CARDIAC PACING IN CHILDREN [J].
ESPERER, HD ;
SINGER, H ;
RIEDE, FT ;
BLUM, U ;
MAHMOUD, FO ;
WENIGER, J .
THORACIC AND CARDIOVASCULAR SURGEON, 1993, 41 (01) :21-27
[5]   Risk factors for venous obstruction in children with transvenous pacing leads [J].
Figa, FH ;
McCrindle, BW ;
Bigras, JL ;
Hamilton, RM ;
Gow, RM .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (08) :1902-1909
[6]   PEDIATRIC TRANSVENOUS PACING - A CONCERN FOR VENOUS THROMBOSIS [J].
GILLETTE, PC ;
ZEIGLER, V ;
BRADHAM, GB ;
KINSELLA, P .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (11) :1935-1939
[7]   ACC/AHA Guidelines for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices: Executive summary - A report of the American College of Cardiology American Heart Association Task Force on Practice Guidelines (Committee on Pacemaker Implantation) [J].
Gregoratos, G ;
Cheitlin, MD ;
Conill, A ;
Epstein, AE ;
Fellows, C ;
Ferguson, TB ;
Freedman, RA ;
Hlatky, MA ;
Naccarelli, GV ;
Saksena, S ;
Schlant, RC ;
Silka, MJ .
CIRCULATION, 1998, 97 (13) :1325-1335
[8]   STEROID-ELUTING EPICARDIAL LEADS IN PEDIATRICS - IMPROVED EPICARDIAL THRESHOLDS IN THE 1ST YEAR [J].
HAMILTON, R ;
GOW, R ;
BAHORIC, B ;
GRIFFITHS, J ;
FREEDOM, R ;
WILLIAMS, W .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (11) :2066-2072
[9]   STEROID-ELUTING EPICARDIAL PACING LEADS IN PEDIATRIC-PATIENTS - ENCOURAGING EARLY RESULTS [J].
JOHNS, JA ;
FISH, FA ;
BURGER, JD ;
HAMMON, JW .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 20 (02) :395-401
[10]   In-line bipolar, steroid-eluting, high impedance, epimyocardial pacing lead [J].
Karpawich, PP ;
Stokes, KB ;
Proctor, K ;
Schallhorn, R ;
McVenes, R .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (03) :503-508