Wear simulation of total hip prostheses with polyethylene against CoCr, alumina and diamond-like carbon

被引:132
作者
Saikko, V
Ahlroos, T
Calonius, O
Keränen, J
机构
[1] Helsinki Univ Technol, Dept Mech Engn, FIN-02015 HUT, Finland
[2] Tampere Univ Technol, Dept Mat Engn, FIN-33101 Tampere, Finland
基金
芬兰科学院;
关键词
total hip arthroplasty; wear; ultra-high molecular weight polyethylene; cobalt-chrome; alumina; diamond-like carbon;
D O I
10.1016/S0142-9612(00)00306-9
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
The wear of ultra-high molecular weight polyethylene acetabular cups was studied with a new biaxial hip wear simulator using diluted calf serum as a lubricant. The cups had been packed and gamma-irradiated in argon. The cups were articulated against two established types of femoral head, alumina and CoCr, and one experimental type, CoCr coated with diamond-like carbon (DLC). The diameter of the heads was 28 mm. Polyethylene against alumina and against CoCr were studied because their clinical wear behaviour is relatively well known. The new simulator was validated with these established materials. The wear mechanisms, including the size and shape of the wear particles, agreed well with those seen in clinical studies. The average wear rates of the cups against alumina and CoCr heads were 48 and 56 mg per 1 million cycles, respectively. The order is in agreement with clinical observations. The average wear rate against DLC was 58 mg per 1 million cycles. As a counterface for polyethylene, DLC did not markedly differ from alumina and CoCr, (C) 2001 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:1507 / 1514
页数:8
相关论文
共 28 条
[1]
HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1993, 26 (08) :969-990
[2]
Wear of ultra-high molecular weight polyethylene acetabular cups in a physiological hip joint simulator in the anatomical position using bovine serum as a lubricant [J].
Bigsby, RJA ;
Hardaker, CS ;
Fisher, J .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1997, 211 (03) :265-269
[3]
Bragdon C R, 1996, Proc Inst Mech Eng H, V210, P157, DOI 10.1243/PIME_PROC_1996_210_408_02
[4]
ISOLATION OF PREDOMINANTLY SUBMICRON-SIZED UHMWPE WEAR PARTICLES FROM PERIPROSTHETIC TISSUES [J].
CAMPBELL, P ;
MA, S ;
YEOM, B ;
MCKELLOP, H ;
SCHMALZRIED, TP ;
AMSTUTZ, HC .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1995, 29 (01) :127-131
[5]
Charnley wear model for validation of hip simulators - Ball diameter versus polytetrafluoroethylene and polyethylene wear [J].
Clarke, IC ;
Good, V ;
Anissian, L ;
Gustafson, A .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1997, 211 (01) :25-36
[6]
THE LETHAL ACTION OF AMINOGLYCOSIDES [J].
DAVIS, BD .
JOURNAL OF ANTIMICROBIAL CHEMOTHERAPY, 1988, 22 (01) :1-3
[7]
Devane PA, 1999, CLIN ORTHOP RELAT R, P59
[8]
DEVANE PA, 1995, CLIN ORTHOP RELAT R, P317
[9]
CHARACTERISTICS OF ACETABULAR CUPS WORN IN HUMAN-BODY [J].
DOWLING, JM ;
ATKINSON, JR ;
DOWSON, D ;
CHARNLEY, J .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1978, 60 (03) :375-382
[10]
Hall R M, 1996, Proc Inst Mech Eng H, V210, P197, DOI 10.1243/PIME_PROC_1996_210_413_02