Combined ultrasound and optoacoustic system for real-time high-contrast vascular imaging in vivo

被引:295
作者
Niederhauser, JJ
Jaeger, M
Lemor, R
Weber, P
Frenz, M [1 ]
机构
[1] Univ Bern, Inst Appl Phys, CH-3012 Bern, Switzerland
[2] Fraunhofer Inst Biomed Engn, D-66386 St Ingbert, Germany
关键词
imaging; in vivo; laser; optoacoustic; photo-acoustic; ultrasound; vascular;
D O I
10.1109/TMI.2004.843199
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
In optoacoustic imaging, short laser pulses irradiate highly scattering human tissue and adiabatically heat embedded absorbing structures, such as blood vessels, to generate ultrasound transients by means of the thermoelastic effect. We present an optoacoustic vascular imaging system that records these transients on the skin surface with an ultrasound transducer array and displays the images online. With a single laser pulse a complete optoacoustic B-mode image can be acquired. The optoacoustic system exploits the high intrinsic optical contrast of blood and provides high-contrast images without the need for contrast agents. The high spatial resolution of the system is determined by the acoustic propagation and is limited to the submillimeter range by our 7.5-MHz linear array transducer. A Q-switched alexandrite laser emitting short near-infrared laser pulses at a wavelength of 760 nm allows an imaging depth of a few centimeters. The system provides real-time images at frame-rates of 7.5 Hz and optionally displays the classically generated ultrasound image alongside the optoacoustic image. The functionality of the system was demonstrated in vivo on human finger, arm and leg. The proposed system combines the merits and most compelling features of optics and ultrasound in a single high-contrast vascular imaging device.
引用
收藏
页码:436 / 440
页数:5
相关论文
共 17 条
[1]   Laser-generated cavitation in absorbing liquid induced by acoustic diffraction [J].
Frenz, M ;
Paltauf, G ;
SchmidtKloiber, H .
PHYSICAL REVIEW LETTERS, 1996, 76 (19) :3546-3549
[2]   Three-dimensional photoacoustic imaging of blood vessels in tissue [J].
Hoelen, CGA ;
de Mul, FFM ;
Pongers, R ;
Dekker, A .
OPTICS LETTERS, 1998, 23 (08) :648-650
[3]   Backward mode detection of laser-induced wide-band ultrasonic transients with optoacoustic transducer [J].
Karabutov, AA ;
Savateeva, EV ;
Podymova, NB ;
Oraevsky, AA .
JOURNAL OF APPLIED PHYSICS, 2000, 87 (04) :2003-2014
[4]   In vivo photoacoustic imaging of blood vessels using an extreme-narrow aperture sensor [J].
Kolkman, RGM ;
Hondebrink, E ;
Steenbergen, W ;
de Mul, FFM .
IEEE JOURNAL OF SELECTED TOPICS IN QUANTUM ELECTRONICS, 2003, 9 (02) :343-346
[5]   Two-dimensional photoacoustic imaging by use of Fourier-transform image reconstruction and a detector with an anisotropic response [J].
Köstli, KP ;
Beard, PC .
APPLIED OPTICS, 2003, 42 (10) :1899-1908
[6]   Optoacoustic imaging using a three-dimensional reconstruction algorithm [J].
Köstli, KP ;
Frauchiger, D ;
Niederhauser, JJ ;
Paltauf, G ;
Weber, HP ;
Frenz, M .
IEEE JOURNAL OF SELECTED TOPICS IN QUANTUM ELECTRONICS, 2001, 7 (06) :918-923
[7]   PHOTOACOUSTIC ULTRASOUND (PAUS) - RECONSTRUCTION TOMOGRAPHY [J].
KRUGER, RA ;
LIU, PY ;
FANG, YR ;
APPLEDORN, CR .
MEDICAL PHYSICS, 1995, 22 (10) :1605-1609
[8]   Breast cancer in vivo: Contrast enhancement with thermoacoustic CT at 434 MHz - Feasibility study [J].
Kruger, RA ;
Miller, KD ;
Reynolds, HE ;
Kiser, WL ;
Reinecke, DR ;
Kruger, GA .
RADIOLOGY, 2000, 216 (01) :279-283
[9]   Real-time three-dimensional optoacoustic imaging using an acoustic lens system [J].
Niederhauser, JJ ;
Jaeger, M ;
Frenz, M .
APPLIED PHYSICS LETTERS, 2004, 85 (05) :846-848
[10]   Comparison of laser-induced and classical ultrasound [J].
Niederhauser, JJ ;
Jaeger, M ;
Frenz, M .
BIOMEDICAL OPTOACOUSTICS IV, 2003, 4960 :118-123