Retrospective motion Gating in small animal CT of mice and rats

被引:53
作者
Bartling, Soenke H.
Stiller, Wolfram
Grasruck, Michael
Schmidt, Bernhard
Peschke, Peter
Semmler, Wolfhard
Kiessling, Fabian
机构
[1] German Canc Res Ctr, Dept Med Phys Radiol, Junior Grp Mol Imaging, D-69120 Heidelberg, Germany
[2] German Canc Res Ctr, Junior Grp Mol Imaging, D-6900 Heidelberg, Germany
[3] German Canc Res Ctr, Dept Med Phys Radiol, D-6900 Heidelberg, Germany
[4] German Canc Res Ctr, Clin Cooperat Unit Radiotherapy, D-6900 Heidelberg, Germany
[5] Siemens Med Solut, Forchheim, Germany
关键词
CT; small animal imaging; flat-panel detector; retrospective; gating;
D O I
10.1097/RLI.0b013e318070dcad
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: Implementation and evaluation of retrospective respiratory and cardiac gating of mice and rats using a flat-panel volume-CT prototype (fpVCT). Materials and Methods: Respiratory and cardiac gating was implemented by equipping a fpVCT with a small animal monitoring unit. ECG and breathing excursions were recorded and 2 binary gating signals derived. Mice and rats were scanned continuously over 80 seconds after administration of blood-pool contrast media. Projections were chosen to reconstruct volumes that fall within defined phases of the cardiac/respiratory cycle. Results: Multireader analysis indicated that in gated still images motion artifacts were strongly reduced and diaphragm, tracheobronchial tract, heart, and vessels sharply delineated. From 4D series, functional data such as respiratory tidal volume and cardiac ejection fraction were calculated and matched well with values known from literature. Discussion: Implementation of retrospective gating in tpVCT improves image quality and opens new perspectives for functional cardiac and lung imaging in small animals.
引用
收藏
页码:704 / 714
页数:11
相关论文
共 36 条
[1]   Micro-CT with respiratory and cardiac gating [J].
Badea, C ;
Hedlund, LW ;
Johnson, GA .
MEDICAL PHYSICS, 2004, 31 (12) :3324-3329
[2]  
Badea C.T., 2005, Molecular Imaging, V4, DOI [10.1162/15353500200504187, DOI 10.1162/15353500200504187]
[3]   Imaging methods for morphological and functional phenotyping of the rodent heart [J].
Badea, CT ;
Bucholz, E ;
Hedlund, LW ;
Rockman, HA ;
Johnson, GA .
TOXICOLOGIC PATHOLOGY, 2006, 34 (01) :111-117
[4]   Polyiodinated triglyceride lipid emulsions for use as hepatoselective contrast agents in CT - Effects of physicochemical properties on biodistribution and imaging profiles [J].
Bakan, DA ;
Weichert, JP ;
Longino, MA ;
Counsell, RE .
INVESTIGATIVE RADIOLOGY, 2000, 35 (03) :158-169
[5]  
BARTLING S, IN PRESS CURR MED IM
[6]   Murine lung tumor measurement using respiratory-gated micro-computed tomography [J].
Cody, DD ;
Nelson, CL ;
Bradley, WM ;
Wislez, M ;
Juroske, D ;
Price, RE ;
Zhou, X ;
Bekele, BN ;
Kurie, JM .
INVESTIGATIVE RADIOLOGY, 2005, 40 (05) :263-269
[7]   Gender and strain variations in left ventricular cardiac function and mass determined with magnetic resonance imaging at 7 tesla in adult mice [J].
Croisille, Pierre ;
Rotaru, Carmen ;
Janier, Marc ;
Hiba, Bassem .
INVESTIGATIVE RADIOLOGY, 2007, 42 (01) :1-7
[8]   High-resolution X-ray microtomography for the detection of lung tumors in living mice [J].
De Clerck, NM ;
Meurrens, K ;
Weiler, H ;
Van Dyck, D ;
Vanhoutte, G ;
Terpstra, P ;
Postnov, AA .
NEOPLASIA, 2004, 6 (04) :374-379
[9]   Fast retrospectively gated quantitative four-dimensional (4D) cardiac micro computed tomography imaging of free-breathing mice [J].
Drangova, Maria ;
Ford, Nancy L. ;
Detombe, Sarah A. ;
Wheatley, Andrew R. ;
Holdsworth, David W. .
INVESTIGATIVE RADIOLOGY, 2007, 42 (02) :85-94
[10]   PRACTICAL CONE-BEAM ALGORITHM [J].