Does ion release differ between hip resurfacing and metal-on-metal THA?

被引:47
作者
Moroni, Antonio [2 ]
Savarino, Lucia [1 ]
Cadossi, Matteo [2 ]
Baldini, Nicola [1 ]
Giannini, Sandro [2 ]
机构
[1] Ist Ortoped Rizzoli, Lab Pathophysiol Orthopaed Implants, I-40136 Bologna, Italy
[2] Ist Ortoped Rizzoli, Div 6, I-40136 Bologna, Italy
关键词
D O I
10.1007/s11999-007-0106-2
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Modern metal-on-metal hip resurfacing was introduced as a bone-preserving method of joint reconstruction for young and active patients; however, the large diameter of the bearing surfaces is of concern for potential increased metal ion release. We hypothesized there were no differences in serum concentrations of chromium, cobalt, and molybdenum between patients who had metal-on-metal hip resurfacing (Group A; average head diameter, 48 mm; median followup, 24 months) and patients who had 28-mm metal-on-metal THA (Group B; median followup, 25 months). Serum concentrations also were compared with concentrations in healthy subjects. We identified no differences in ion levels between Groups A and B. A distinction was made according to gender. Women showed a higher chromium release in Group A whereas men had a higher cobalt release in Group B. Values obtained from Group A were higher than those of the control subjects. Our data suggest metal-on-metal bearings for THA should not be rejected because of concern regarding potential increased metal ion release; however, patients with elevated ion levels, even without loosening or toxicity, could be at higher risk and should be followed up periodically.
引用
收藏
页码:700 / 707
页数:8
相关论文
共 47 条
[1]
How do serum cobalt and chromium levels change after metal-on-metal hip resurfacing? [J].
Back, DL ;
Young, AA ;
Shimmin, AJ .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (438) :177-181
[2]
Serum cobalt levels after metal-on-metal total hip arthroplasty [J].
Brodner, W ;
Bitzan, P ;
Meisinger, V ;
Kaider, A ;
Gottsauner-Wolf, F ;
Kotz, R .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (11) :2168-2173
[3]
Cup inclination and serum concentration of cobalt and chromium after metal-on-metal total hip arthroplasty [J].
Brodner, W ;
Grübl, A ;
Jankovsky, R ;
Meisinger, V ;
Lehr, S ;
Gottsauner-Wolf, F .
JOURNAL OF ARTHROPLASTY, 2004, 19 (08) :66-70
[4]
BUTCHER DJ, 1998, SERIES MONOGRAPHS AN
[5]
Levels of metal ions after small- and large-diameter metal-on-metal hip arthroplasty [J].
Clarke, MT ;
Lee, PTH ;
Arora, A ;
Villar, RN .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2003, 85B (06) :913-917
[6]
The effect of the diameter of metal-on-metal bearings on systemic exposure to cobalt and chromium [J].
Daniel, J ;
Ziaee, H ;
Salama, A ;
Pradhan, C ;
McMinn, DJW .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2006, 88B (04) :443-448
[7]
Range of motion after total hip resurfacing [J].
dela Rosa, Mylene A. ;
Silva, Mauricio ;
Heisel, Christian ;
Reich, Michael ;
Schmalzried, Thomas P. .
ORTHOPEDICS, 2007, 30 (05) :352-357
[8]
Doorn PF, 1998, J BIOMED MATER RES, V42, P103, DOI 10.1002/(SICI)1097-4636(199810)42:1<103::AID-JBM13>3.3.CO
[9]
2-D
[10]
Dorr L D, 1994, J Arthroplasty, V9, P453, DOI 10.1016/0883-5403(94)90090-6