Amplified, frequency swept lasers for frequency domain reflectometry and OCT imaging: design and scaling principles

被引:389
作者
Huber, R [1 ]
Wojtkowski, M
Taira, K
Fujimoto, JG
Hsu, K
机构
[1] MIT, Dept Elect Engn & Comp Sci, Cambridge, MA 02139 USA
[2] MIT, Elect Res Lab, Cambridge, MA 02139 USA
[3] Micron Opt Inc, Atlanta, GA 30345 USA
关键词
D O I
10.1364/OPEX.13.003513
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We demonstrate a high-speed, frequency swept, 1300 nm laser source for frequency domain reflectometry and OCT with Fourier domain/swept-source detection. The laser uses a fiber coupled, semiconductor amplifier and a tunable fiber Fabry-Perot filter. We present scaling principles which predict the maximum frequency sweep speed and trade offs in output power, noise and instantaneous linewidth performance. The use of an amplification stage for increasing output power and for spectral shaping is discussed in detail. The laser generates similar to 45 mW instantaneous peak power at 20 kHz sweep rates with a tuning range of similar to 120 nm full width. In frequency domain reflectometry and OCT applications the frequency swept laser achieves 108 dB sensitivity and similar to 10 mu m axial resolution in tissue. We also present a fast algorithm for real time calibration of the fringe signal to equally spaced sampling in frequency for high speed OCT image preview. (C) 2005 Optical Society of America.
引用
收藏
页码:3513 / 3528
页数:16
相关论文
共 19 条
[1]   MODIFIED OPTICAL FREQUENCY-DOMAIN REFLECTOMETRY WITH HIGH SPATIAL-RESOLUTION FOR COMPONENTS OF INTEGRATED OPTIC SYSTEMS [J].
BARFUSS, H ;
BRINKMEYER, E .
JOURNAL OF LIGHTWAVE TECHNOLOGY, 1989, 7 (01) :3-10
[2]   Optical coherence tomography using a frequency-tunable optical source [J].
Chinn, SR ;
Swanson, EA ;
Fujimoto, JG .
OPTICS LETTERS, 1997, 22 (05) :340-342
[3]   Sensitivity advantage of swept source and Fourier domain optical coherence tomography [J].
Choma, MA ;
Sarunic, MV ;
Yang, CH ;
Izatt, JA .
OPTICS EXPRESS, 2003, 11 (18) :2183-2189
[4]   Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography [J].
de Boer, JF ;
Cense, B ;
Park, BH ;
Pierce, MC ;
Tearney, GJ ;
Bouma, BE .
OPTICS LETTERS, 2003, 28 (21) :2067-2069
[5]   COHERENT FREQUENCY-DOMAIN REFLECTOMETRY FOR CHARACTERIZATION OF SINGLE-MODE INTEGRATED-OPTICAL WAVE-GUIDES [J].
GLOMBITZA, U ;
BRINKMEYER, E .
JOURNAL OF LIGHTWAVE TECHNOLOGY, 1993, 11 (08) :1377-1384
[6]   Optical frequency-domain reflectometry using rapid wavelength tuning of a Cr4+:forsterite laser [J].
Golubovic, B ;
Bouma, BE ;
Tearney, GJ ;
Fujimoto, JG .
OPTICS LETTERS, 1997, 22 (22) :1704-1706
[7]   OPTICAL COHERENCE TOMOGRAPHY [J].
HUANG, D ;
SWANSON, EA ;
LIN, CP ;
SCHUMAN, JS ;
STINSON, WG ;
CHANG, W ;
HEE, MR ;
FLOTTE, T ;
GREGORY, K ;
PULIAFITO, CA ;
FUJIMOTO, JG .
SCIENCE, 1991, 254 (5035) :1178-1181
[8]   Linearizing optical frequency-sweep of a laser diode for FMCW reflectometry [J].
Iiyama, K ;
Wang, LT ;
Hayashi, K .
JOURNAL OF LIGHTWAVE TECHNOLOGY, 1996, 14 (02) :173-178
[9]   Performance of fourier domain vs. time domain optical coherence tomography [J].
Leitgeb, R ;
Hitzenberger, CK ;
Fercher, AF .
OPTICS EXPRESS, 2003, 11 (08) :889-894
[10]   INTERFEROMETRIC DETERMINATION OF A STATIC OPTICAL-PATH DIFFERENCE USING A FREQUENCY SWEPT LASER DIODE [J].
OHBA, R ;
UEHIRA, I ;
KAKUMA, S .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1990, 1 (06) :500-504